Skip to the content

FDA Medical Device Recalls (openFDA)

Kelyniam Global, Inc.: Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1002

Recall number
Z-1871-2018
Recalling firm
Kelyniam Global, Inc.
Product
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1002
Status
Terminated
Initiated
2017-06-10
Posted
not on file
Terminated
2020-09-14
Reason
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
Root cause
Device Design
Action
On June 10, 2017, the recalling firm notified its Sales Representatives that effective June 10, 2017, the firm would no longer offer the temporal suture system (TSS). The firm requested that sales reps dispose of all marketing materials which contain details regarding the TSS. Sales reps who wanted to keep existing booklets in the field were asked to contact the firm to obtain adhesive labels to place over the discontinued product. On November 13, 2017, the recalling firm notified its Sales Representatives that the firm was temporarily suspending shipment of the Integrated Fixation System (IFS). The firm stated that all paper brochures were to be discarded, and asked sales reps to inform all those who received "Customize Skull and Craniofacial Implants" brochure and IFU. Revised marketing materials are available on the firm's website, www.kelyniam.com. If you have any questions, you may contact Dr. Mark V. Smith, Vice President of Kelyniam Global, Inc., at msmith@kelyniam.com, or by phone at (800) 280-8192 x702.
Quantity
16
Distribution
US nationwide distribution.
Lot, serial or model codes
Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C.  Case Numbers: CSI082114-CT1 CSI081214-TB1 CSI020216-LK1 CSI092016-LD1R CSI100116-BM1A CSI092616-YB1 CSI092616-MM1 CSI112316-YB1 CSI050217-LS1 CSI052217-AF1 CSI062317-GM1 CSI071017-NS1B CSI071417-CM1 CSI102617-DL1 CSI101617-JF1A CSI102617-DL1
510(k) numbers
["K103582"]
PMA numbers
not on file
Product code
GXN
Firm FEI
3009424796
Firm city
Collinsville
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register