FDA Medical Device Recalls (openFDA)
Intuitive Surgical, Inc.: IS4000 da Vinci Xi - Endoscope accessory; 0 Degree, 8mm Endoscope. Intuitive Surgical Inc.
- Recall number
- Z-1871-2014
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- IS4000 da Vinci Xi - Endoscope accessory; 0 Degree, 8mm Endoscope. Intuitive Surgical Inc.
- Status
- Terminated
- Initiated
- 2014-06-11
- Posted
- 2014-06-25
- Terminated
- 2014-07-02
- Reason
- In some procedures when using the endoscope in conjunction with the IS4000 (da Vinci Xi) System, the surgical view has an orange-yellow tint which can make it hard to differentiate between tissue
- Root cause
- Device Design
- Action
- Medical Device Recall notifications were sent to the 4 affected sites on June 11, 2014. Letters describe the problem and the potential health risk. The two model numbers of the endoscope are listed and the statement that ISI will perform a software upgrade to the da Vinci System and replace the endoscopes. Letters request that the following actions take place: - Ensure that all appropriate personnel are fully informed of the information provided in the notification letter. - Complete and return the acknowledgement form using instructions provided. - Retain a copy of The letter for your records. Intuitive representatives will schedule a site visit to provide the re-calibrated endoscope, update the software and ensure that it is properly integrated with the da Vinci Xi Surgical System. Questions or concerns may be directed to Intuitive Customer Service at 800-876-1310.
- Quantity
- 19 units
- Distribution
- Distributed in the states of FL, TX, and NY.
- Lot, serial or model codes
- 470026-18
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as