Intuitive Surgical, Inc.: IS4000 da Vinci Xi - Endoscope accessory; 0 Degree, 8mm Endoscope. Intuitive Surgical Inc.

Recall number
Z-1871-2014
Recalling firm
Intuitive Surgical, Inc.
Product
IS4000 da Vinci Xi - Endoscope accessory; 0 Degree, 8mm Endoscope. Intuitive Surgical Inc.
Status
Terminated
Initiated
2014-06-11
Posted
2014-06-25
Terminated
2014-07-02
Reason
In some procedures when using the endoscope in conjunction with the IS4000 (da Vinci Xi) System, the surgical view has an orange-yellow tint which can make it hard to differentiate between tissue
Root cause
Device Design
Action
Medical Device Recall notifications were sent to the 4 affected sites on June 11, 2014. Letters describe the problem and the potential health risk. The two model numbers of the endoscope are listed and the statement that ISI will perform a software upgrade to the da Vinci System and replace the endoscopes. Letters request that the following actions take place: - Ensure that all appropriate personnel are fully informed of the information provided in the notification letter. - Complete and return the acknowledgement form using instructions provided. - Retain a copy of The letter for your records. Intuitive representatives will schedule a site visit to provide the re-calibrated endoscope, update the software and ensure that it is properly integrated with the da Vinci Xi Surgical System. Questions or concerns may be directed to Intuitive Customer Service at 800-876-1310.
Quantity
19 units
Distribution
Distributed in the states of FL, TX, and NY.
Lot, serial or model codes
470026-18
510(k) numbers
not on file
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register