Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: GE Healthcare, Uni-Circuit Kit, Adult, Disposable, 1.8 m/72 in, REF M1089028, M1087972, M1087975, M1126445, M1127508, M1127509, M1129780, M1129781, M1130230, M1137805, M1137199, M1145542, M1170604, M1151662, M1152224, M1158990, M1100486, M1122080, M1126444, M1163003, M1163255 M1132533, M1133966, M1170051, M1135144, M1139485, M1118627, M1118628, M1145406, M1147298, M1137811, M1152651, M1160275. pkg of 20 pcs, Latex Free. Breathing circuit for use with anesthesia machine or ventilator.

Recall number
Z-1870-2009
Recalling firm
GE Medical Systems, LLC
Product
GE Healthcare, Uni-Circuit Kit, Adult, Disposable, 1.8 m/72 in, REF M1089028, M1087972, M1087975, M1126445, M1127508, M1127509, M1129780, M1129781, M1130230, M1137805, M1137199, M1145542, M1170604, M1151662, M1152224, M1158990, M1100486, M1122080, M1126444, M1163003, M1163255 M1132533, M1133966, M1170051, M1135144, M1139485, M1118627, M1118628, M1145406, M1147298, M1137811, M1152651, M1160275. pkg of 20 pcs, Latex Free. Breathing circuit for use with anesthesia machine or ventilator.
Status
Terminated
Initiated
2009-06-05
Posted
2009-09-09
Terminated
2014-01-16
Reason
Breathing circuit inner tube may separate from the rigid connectors under certain conditions. A separation of the inner tube from the connector may prevent adequate gas flow (O2, N2O and anesthetic agents) to the patient. A separation can also cause CO2 re-breathing. This could lead to hypoxia and/or hypercarbia and inadequate anesthesia delivery.
Root cause
Process control
Action
GE Healthcare sent consignees 6/5/09 an "Urgent Medical Device Correction" letter dated June 5, 2009. The letter was addressed to Health Care Administrator/Risk Manager; Chief of Anesthesia; Director of Clinical Engineering; Risk Manager. The letter described the Safety Issue, Affected Product Details, Safety Instructions (included "Recall Confirmation" form fax to 973-790-4150), Product Correction and Contact Information.
Quantity
85097
Distribution
Worldwide distribution: USA, AUSTRALIA, DENMARK, FRANCE, ESTONIA, GERMANY, ITALY, JORDAN, NETHERLANDS, NORWAY, SWEDEN, and UNITED ARAB EMIRATES.
Lot, serial or model codes
This correction pertains to Uni-Circuits from lot# 084812 and later, up to and including lot# 092210
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAI
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register