Fresenius Medical Care Holdings, Inc.: 2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.

Recall number
Z-1869-2024
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
Status
Open, Classified
Initiated
2024-04-17
Posted
2024-05-17
Terminated
not on file
Reason
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Root cause
Device Design
Action
On April 17, 2024 IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer actions: As per the 2008T Operators Manual P/N 490122 Rev AA: Blood Pump Rotors should be inspected for proper operation, as stated in Chapter 2: Daily preparation for treatment. If you experience a dislodged guide post sleeve, bent or loose tubing guide posts, or roller cannot move freely on your current F40015481 Rev A blood pump rotor, please contact FMCRTG Technical Support Services at 800-405-1321 for a replacement Blood Pump Rotor (Part Number, F40015481, Rev C) . Have Rotor Lot Number available when requesting a replacement. Please provide this letter to your Biomedical Technician or trained personnel. Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatch at https://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. Provide this Customer Notice to all those who need to be aware within your organization. Complete and return the attached reply form. Biomedical Technician/ actions: A Field Service Bulletin will be posted to fmcna.com website regarding Preventative Maintenance activities for Blood Pump Rotor Tubing Guide Inspection of the need to inspect the blood pump tubing guides during the 6-month Preventive Maintenance on all 2008 Series Hemodialysis Machines. URL: https://fmcna.com/support/product-support-documents/field-service-bulletins/ For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309.
Quantity
334 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Component: Model Number: F40015481 Rev A, UDI/DI: N/A, All Rev A lots. Finished Devices: Model Number: 191128, UDI/DI: 00840861102105; Serial Numbers: 3T0S274210, 1T0S260183, 1T0S260455, 1T0S260456, 1T0S260485, 1T0S260273, 1T0S260283, 1T0S260490, 1T0S260158, 1T0S260182, 1T0S260266, 3T0S269011, 3T0S273223, 3T0S274216, 3T0S274225, 3T0S274678, 3T0S274680, 1T0S260281, 2T0S268008, 2T0S268016, 2T0S264775, 2T0S268009, 2T0S268013, 2T0S268014, 2T0S268015, 3T0S269021, 3T0S269026, 3T0S274215, 3T0S274222, 3T0S274231, 3T0S274234, 1T0S260154, 1T0S260481, 1T0S260543, 1T0S260550, 1T0S260552, 3T0S274213, 3T0S274230, 1T0S260242, 1T0S260250, 1T0S260510, 1T0S260514, 2T0S267901, 2T0S267905, 2T0S267906, 2T0S267911, 1T0S260541, 1T0S260546, 2T0S264780, 3T0S274233, 3T0S274685, 3T0S274691, 3T0S274697, 1T0S260159, 1T0S260164, 1T0S260172, 1T0S260212, 1T0S260216, 1T0S260221, 1T0S260269, 1T0S260270, 1T0S260276, 1T0S260292, 1T0S260295, 1T0S260528, 3T0S273221, 3T0S274217, 3T0S274218, 3T0S274219, 3T0S274220, 3T0S274223, 3T0S274226, 3T0S274227, 3T0S274228, 3T0S274232, 3T0S274682, 3T0S274684, 3T0S274690, 3T0S274692, 3T0S274693, 3T0S274698, 1T0S260150, 1T0S260174, 1T0S260192, 1T0S260554, 1T0S260534, 3T0S274214, 3T0S274688, 1T0S260286, 2T0S264786, 3T0S274679, 3T0S274681, 3T0S274683, 3T0S274686, 3T0S274694, 3T0S274695, 3T0S274699, 3T0S274700, 1T0S260544, 1T0S260268, 1T0S260294, 3T0S269020, 3T0S269030, 3T0S269031, 1T0S260227, 2T0S267900, 2T0S267904, 1T0S260284, 1T0S260457, 1T0S260548, 3T0S269013, 3T0S269015, 3T0S269028, 3T0S271557, 3T0S271562, 3T0S271563, 3T0S271570, 3T0S271571, 3T0S271573, 3T0S271574, 3T0S271575, 3T0S271576, 3T0S269033, 3T0S269035, 3T0S271561, 3T0S271566, 3T0S271568, 3T0S273244, 1T0S260493, 1T0S260501, 1T0S260506, 1T0S260515, 1T0S260526, 1T0S260279, 1T0S260155, 1T0S260156, 1T0S260166, 1T0S260167, 1T0S260185, 1T0S260188, 1T0S260200, 3T0S269014, 3T0S269016, 3T0S269022, 2T0S264805, 2T0S264776, 2T0S264778, 2T0S264794, 2T0S264818, 2T0S264822, 2T0S264787, 2T0S264792, 3T0S269017, 3T0S269018, 3T0S269019, 3T0S271558, 3T0S271560, 3T0S271564, 3T0S271565, 3T0S271572, 1T0S256406, 2T0S267910, 3T0S273219, 3T0S273232, 3T0S274221, 1T0S260151, 1T0S260178, 1T0S260152, 1T0S260278, 1T0S260288, 1T0S260488, 1T0S260533, 1T0S260249, 1T0S260259, 1T0S260262, 1T0S260198, 1T0S260291, 2T0S268012, 1T0S260173, 3T0S274211, 3T0S274224, 2T0S268011, 3T0S273220, 3T0S273222, 3T0S273224, 3T0S273225, 3T0S273226, 3T0S273227, 3T0S273228, 3T0S273229, 3T0S273230, 3T0S273233, 3T0S273234, 3T0S273235, 3T0S273236, 3T0S273237, 3T0S273238, 3T0S273239, 3T0S273240, 3T0S273241, 3T0S273242, 3T0S273243, 1T0S260267, 1T0S260275, 3T0S269012, 3T0S269023, 3T0S269024, 3T0S269025, 3T0S269027, 3T0S269029, 3T0S269032, 3T0S269034, 1T0S260204, 1T0S260460, 1T0S260482, 1T0S260513, 1T0S260516, 1T0S260527, 1T0S260540, 1T0S260545, 2T0S267903, 2T0S264781, 3T0S271559, 3T0S271567, 3T0S271569, 1T0S260290, 1T0S260175, 1T0S260282, 1T0S260508, 1T0S260566, 1T0S260285, 2T0S264783, 2T0S264814, 1T0S260157, 1T0S260161, 1T0S260162, 1T0S260163, 1T0S260165, 1T0S260169, 1T0S260176, 1T0S260177, 1T0S260179, 1T0S260187, 1T0S260189, 1T0S260191, 1T0S260193, 1T0S260195, 1T0S260196, 1T0S260197, 1T0S260199, 1T0S260201, 1T0S260202, 1T0S260203, 1T0S260207, 1T0S260208, 1T0S260209, 1T0S260210, 1T0S260211, 1T0S260213, 1T0S260214, 1T0S260215, 1T0S260217, 1T0S260218, 1T0S260219, 1T0S260220, 1T0S260222, 1T0S260223, 1T0S260224, 1T0S260225, 1T0S260228, 1T0S260229, 1T0S260230, 1T0S260231, 1T0S260232, 1T0S260233, 1T0S260235, 1T0S260236, 1T0S260237, 1T0S260239, 1T0S260243, 1T0S260245, 1T0S260247, 1T0S260251, 1T0S260253, 1T0S260255, 1T0S260256, 1T0S260257, 1T0S260258, 1T0S260260, 1T0S260272, 1T0S260453, 1T0S260458, 1T0S260459, 1T0S260483, 1T0S260500, 1T0S260502, 1T0S260507, 1T0S260519, 1T0S260523, 1T0S260531, 1T0S260542, 1T0S260549, 1T0S260553, 1T0S260565, 2T0S264774, 2T0S264777, 2T0S264779, 2T0S264782, 2T0S264784, 2T0S264788, 2T0S264791, 2T0S264793, 2T0S264796, 2T0S264797, 2T0S264800, 2T0S264801, 2T0S264803, 2T0S264804, 2T0S264807, 2T0S264811, 2T0S264819, 2T0S264821, 2T0S268010, 1T0S260194, 1T0S260238, 1T0S260293, 1T0S260461, 2T0S264802, 3T0S274212, 2T0S267899, 2T0S267902, 2T0S267907, 2T0S267908, 2T0S267909.
510(k) numbers
["K173972","K222952","K231125"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register