Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized fashion within the case, 8mm and 9mm pedicle screws used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients

Recall number
Z-1869-2013
Recalling firm
SpineFrontier, Inc.
Product
Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized fashion within the case, 8mm and 9mm pedicle screws used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients
Status
Terminated
Initiated
2012-10-05
Posted
2013-08-01
Terminated
2015-07-29
Reason
Two sections that are designated for the 8mm and 9mm screws were switched; the holes that hold the 8mm screws are designed for 9mm screws, and the holes that hold the 9mm screws are designated for 8mm screws.
Root cause
Device Design
Action
The firm, SpineFrontier, sent an Advisory Notice entitled "URGENT May 1, 2013, PedFuse Screw Caddy Removal" dated May 1, 2013 to their consignees/customers. The notice described the product, problem and action to be taken. Consignees/customers were informed that a removal has been initiated to remove the parts from the field to ensure that all sets contain correctly marked screw caddies. Consignees/customers wre instructed to complete and return the URGENT PedFuse Screw Caddy Removal ACKNOWLEDGEMENT FORM via fax to: SpineFrontier, Inc., 978-232-3991 or email: customerservice@spinefrontier.com. Replacement caddies will be shipped upon receipt of completed acknowledgement form and RMA request for part return. Please feel free to contact SpineFrontier at 978-232-3990 with any questions.
Quantity
11 caddies
Distribution
US distribution in states of: CO, DC, MA and TX.
Lot, serial or model codes
31176, 018652, 018645R, 42011
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSM
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register