LAP of America Laser Applications, LLC: LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.

Recall number
Z-1866-2014
Recalling firm
LAP of America Laser Applications, LLC
Product
LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.
Status
Terminated
Initiated
2014-03-21
Posted
2014-06-24
Terminated
2016-09-26
Reason
While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.
Root cause
Software design
Action
LAP of America Laser sent an Urgent safety Information FFST 2014-2-3 via UPS Next Day Air to the customer.
Quantity
1 system
Distribution
North Carolina
Lot, serial or model codes
Product Code: 0007569-0001
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
1000113153
Firm city
Boynton Beach
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register