Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Ultraflex Esophageal Stent System. Uncovered - Proximal Release - 18/23/10. Catalog Number M00513810. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Intended for maintaining esophageal luminal patency in esophageal structures caused by intrinsic and/or extrinsic malignant tumors only and occlusion of concurrent esophageal fistulas.

Recall number
Z-1865-2010
Recalling firm
Boston Scientific Corporation
Product
Ultraflex Esophageal Stent System. Uncovered - Proximal Release - 18/23/10. Catalog Number M00513810. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Intended for maintaining esophageal luminal patency in esophageal structures caused by intrinsic and/or extrinsic malignant tumors only and occlusion of concurrent esophageal fistulas.
Status
Terminated
Initiated
2010-05-17
Posted
2010-06-25
Terminated
2011-10-20
Reason
Suture related device failures during deployment and repositioning of the stent
Root cause
Device Design
Action
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Quantity
not on file
Distribution
Worldwide Distribution - Only US distributed product is affected by recall
Lot, serial or model codes
Lot Numbers: 11720486 11741834 11750686 11788787 11804318 11814051 11824077 11846108 11870368 11870594 11909641 11931304 11974372 11982929 12001740 12019957 12074676 12157485 12176238 12184464 12229977 12240225 12297673 12305981 12406877 12464947 12536118 12571835 12882958 12914130 12931878
510(k) numbers
["K032930"]
PMA numbers
not on file
Product code
ESW
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register