NxStage Medical Inc: NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.

Recall number
Z-1864-2025
Recalling firm
NxStage Medical Inc
Product
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
Status
Open, Classified
Initiated
2025-04-28
Posted
2025-05-28
Terminated
not on file
Reason
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
Root cause
Under Investigation by firm
Action
NxStage notified consignees via mail on 04/28/2025. Consignees were instructed to check inventory for affected units and cease use if any are on hand, complete and return the response form, and contact NxStage Customer Service to arrange return of all affected units.
Quantity
115,465 cases (230,930 units)
Distribution
US Nationwide distribution.
Lot, serial or model codes
a. UDI-DI number: M535RFP4000. Device Catalog Number RFP-400. Lot Numbers: Q2407307, Q2407308, Q2407309. b. UDI-DI number: M535RFP4010. Device Catalog Number RFP-401. Lot Numbers: Q2407061, Q2407147, Q2407148, Q2407149, Q2407313, Q2407386, Q2407387, Q2407388, Q2407389, Q2407390. c. UDI-DI number: M535RFP4020. Device Catalog Number RFP-402. Lot Numbers: Q2407213, Q2407214. d. UDI-DI number: M535RFP4030. Device Catalog Number RFP-403. Lot Number: Q2407311. e. UDI-DI number: M535RFP4040. Device Catalog Number RFP-404. Lot Numbers: Q2407062, Q2407063, Q2407064, Q2407065, Q2407067, Q2407068, Q2407069, Q2407070, Q2407071, Q2407217, Q2407218, Q2407219, Q2407220, Q2407221, Q2407222, Q2407223, Q2407224, Q2407225, Q2407301, Q2407302, Q2407303, Q2407304, Q2407305, Q2407306, Q2407391, Q2407392, Q2407405, Q2407406. f. UDI-DI number: M535RFP4060. Device Catalog Number RFP-406. Lot Numbers: Q2407072, Q2407137, Q2407140, Q2407141, Q2407216, Q2407310, Q2407312. g. UDI-DI number: M535RFP4530. Device Catalog Number RFP-453. Lot Number: Q2407066. h. UDI-DI number: M535RFP4540. Device Catalog Number RFP-454. Lot Numbers: Q2407145, Q2407146, Q2407226. j. UDI-DI number: M535RFP4560. Device Catalog Number RFP-456. Lot Numbers: Q2407138, Q2407139, Q2407142, Q2407143, Q2407144, Q2407227.
510(k) numbers
["K053286"]
PMA numbers
not on file
Product code
KPO
Firm FEI
3003464075
Firm city
Lawrence
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register