GE Healthcare, LLC: GE Healthcare Innova IGS 3 (GTIN 00840682147378), Innova IGS 5 (GTIN 00840682124621), Innova IGS 6 (GTIN 00840682124614), Discovery IGS 7 (GTIN 00840682124638), and Discovery IGS 7 OR (GTIN00840682125888) interventional fluoroscopic x-ray systems.
- Recall number
- Z-1864-2021
- Recalling firm
- GE Healthcare, LLC
- Product
- GE Healthcare Innova IGS 3 (GTIN 00840682147378), Innova IGS 5 (GTIN 00840682124621), Innova IGS 6 (GTIN 00840682124614), Discovery IGS 7 (GTIN 00840682124638), and Discovery IGS 7 OR (GTIN00840682125888) interventional fluoroscopic x-ray systems.
- Status
- Terminated
- Initiated
- 2021-04-19
- Posted
- 2021-06-15
- Terminated
- 2023-08-31
- Reason
- The IGS system can experience a single vertical line defect where the vertical line divides and horizontally shifts live monitor images into two unequal image parts within the monitor display screen.
- Root cause
- Component change control
- Action
- The recalling firm issued letters dated 4/19/2021 via traceable mail on 4/20/21 and issued an additional letter dated 4/28/2021 to correct an omission of serial numbers (only sent to the affected consignees).
- Quantity
- 139 devices
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, GA, KY, LA, MA, MO, MS, NC, NJ, NV, NY, PA, SC, TN, TX, VA, WA, and WI. Government distribution was made and potentially military distribution was made. The countries of Algeria, Argentina, Canada, China, Dominican Republic, Finland, France, Germany, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Korea, Kuwait, Mexico, Moldova, Morocco, Poland, Russia, Spain, Sudan, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uzbekistan.
- Lot, serial or model codes
- Affected serial numbers: B2-20-006, B2-20-007, B2-20-008, B2-20-009, B2-21-001, B3-20-024, B3-20-026, B3-20-027, B3-20-028, B3-21-001, B3-21-002, B3-21-003, B3-21-006, D3-20-025, D3-20-028, D3-20-029, D3-21-002, D3-21-005, D3-21-006, D3-21-007, D3-21-008, D3-21-009, D4-20-025, D4-20-026, D4-20-029, D4-20-031, D4-20-032, D4-20-033, D4-21-003, D4-21-004, D4-21-005, D4-21-007, D4-21-008, D4-21-010, M2-20-040, M2-20-041, M2-20-042, M2-20-043, M2-20-045, M2-20-046, M2-20-047, M2-20-048, M2-20-049, M2-20-050, M2-20-051, M2-20-052, M2-20-053, M2-20-054, M2-21-003, M2-21-004, M2-21-005, M2-21-006, M2-21-007, M2-21-008, M2-21-009, M2-21-010, M2-21-012, M2-21-013, M2-21-014, M3-20-094, M3-20-095, M3-20-096, M3-20-098, M3-20-101, M3-20-105, M3-20-108, M3-20-110, M3-20-114, M3-20-116, M3-20-118, M3-20-119, M3-20-120, M3-20-122, M3-20-123, M3-20-124, M3-20-125, M3-20-126, M3-20-127, M3-21-001, M3-21-006, M3-21-007, M3-21-009, M3-21-010, M3-21-011, M3-21-012, M3-21-013, M3-21-014, M3-21-015, M3-21-016, M3-21-017, M3-21-019, M3-21-020, M3-21-023, M3-21-024, M3-21-026, M3-21-027, M3-21-028, M4-20-022, M4-20-023, M4-20-026, M4-20-028, M4-20-029, M4-20-030, M4-20-033, M4-20-034, M4-20-035, M4-20-036, M4-20-039, M4-20-041, M4-21-001, M4-21-002, M4-21-004, M4-21-005, M4-21-006, M4-21-007, M4-21-008, M4-21-009, M4-21-011, M4-21-012, M4-21-013, M4-21-014, M4-21-015, M4-21-016, and M4-21-017. Serial numbers that are under GE control but are consigned to a facility and not yet installed: D4-21-009, M3-21-022, M3-21-008, B3-21-005, D3-21-003, M3-21-025, M2-20-044, M3-21-029 D3-20-027, M2-21-011, and M3-21-021.
- 510(k) numbers
- ["K181403"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28