Skip to the content

FDA Medical Device Recalls (openFDA)

Hill-Rom, Inc.: Advanta 2 Bed. The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the hospital.

Recall number
Z-1864-2012
Recalling firm
Hill-Rom, Inc.
Product
Advanta 2 Bed. The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the hospital.
Status
Terminated
Initiated
2012-06-18
Posted
2012-06-22
Terminated
2014-06-06
Reason
During an evaluation of Advanta 2 siderail samples returned from the field it was found that the screws used to attach the siderail were being stripped from the plastic cavity of the siderail.
Root cause
Process design
Action
Hill-Rom sent an "URGENT FIELD SAFETY NOTICE" dated May 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included a Customer Response Form for customers to complete and return to the firm. Contact Hill-Rom Technical Support at 800-445-3720 for questions regarding this notice.
Quantity
7843 devices
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, France, and Singapore.
Lot, serial or model codes
AdvantaTM 2 beds (P1190) produced between August 10, 2009 and October 6, 2011; bed S/N between K222AT9180 and M279AT4403
510(k) numbers
not on file
PMA numbers
not on file
Product code
FNL
Firm FEI
1824206
Firm city
Batesville
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register