FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee Instrumentation system. It interfaces with the Modular Handled and the Femoral Stylus to facilitate sizing for a femoral component, and preparing holes to accept the corresponding 4:1 Cutting Guide. The femoral sizer is designed to prepare the femoral cuts in 3 degrees of external rotation by referencing against the posterior condyles. The Femoral sizer is set by locking it into the left or right orientation based on the femur.

Recall number
Z-1864-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee Instrumentation system. It interfaces with the Modular Handled and the Femoral Stylus to facilitate sizing for a femoral component, and preparing holes to accept the corresponding 4:1 Cutting Guide. The femoral sizer is designed to prepare the femoral cuts in 3 degrees of external rotation by referencing against the posterior condyles. The Femoral sizer is set by locking it into the left or right orientation based on the femur.
Status
Terminated
Initiated
2005-05-27
Posted
2008-09-16
Terminated
2008-09-24
Reason
Stryker Orthopaedics became aware that when attempting to position the Triathlon Femoral Sizer, the pre-setting may slip from l (left knee) to R (right knee) and vice verse intraoperatively.
Root cause
Device Design
Action
Letters were sent to all affected Stryker Branches/Agencies on May 27, 2005. Letters requested that each branch examine the inventory and hospitals locations to identify the instruments. Contact James Young at 201-831-5110, Brian Coughlin at 210-831-5665 or Rita Intorrella at 201-831-5825 if you have questions.
Quantity
560 units
Distribution
To Stryker branches/agencies worldwide.
Lot, serial or model codes
Catalog Number: 6541-1-603 All lots (ecxept lot codes NMZC01, NMYV01, NMYW03.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register