Straumann USA LLC: GM Helix Acqua Implant, Article Number: 140.985
- Recall number
- Z-1863-2025
- Recalling firm
- Straumann USA LLC
- Product
- GM Helix Acqua Implant, Article Number: 140.985
- Status
- Open, Classified
- Initiated
- 2025-05-01
- Posted
- 2025-05-28
- Terminated
- not on file
- Reason
- It is possible that a package labelled as an 11.5 mm implant may contain a 13 mm implant instead.
- Root cause
- Under Investigation by firm
- Action
- On May 1, 2025 URGENT FIELD SAFETY ACTION letters were sent to customers. Action to be taken: 1. If article 140.984 Lot JPZ21 or 109.985 Lot JPZ75 are still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If the product was used and no complications or problems were identified, of the patient did not present symptoms there is no need for any additional action with the patient. 3. If article 140.984 Lot JPZ21 or 109.985 Lot JPZ75 are still in your inventory, return it for a credit or replacement as indicated on the Customer Confirmation Form. 4. In the product is not found in your stock, indicate this on the enclosed Customer Confirmation Form. 5 In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC; 60 Minuteman Road, Andover, MA 01810. Transmission of the Field Safety Notice: This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action. We apologize for any inconvenience that this may cause.
- Quantity
- 207 units (54 US, 153 OUS)
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, Il KS, LA, MA, MD, MI, NE, NH, NJ, NY, OH, OK, RI, SC, TN, TX, UT, WI and the countries of Australia, Croatia, France, Hungary, India, Italy, Lebanon, Lithuania, New Zealand, Portugal, Qatar, Serbia, Spain, Tunisia, UAE, United Kingdom, Vietnam.
- Lot, serial or model codes
- Article Number: 140.985; UDI-DI: (01)7899878024569(11)220125(17)270124(10)JPZ75; Lot number: JPZ75;
- 510(k) numbers
- ["K163194"]
- PMA numbers
- not on file
- Product code
- DZE
- Firm FEI
- 1000121052
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28