FDA Medical Device Recalls (openFDA)
Ultra Clean Systems, Inc.: Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system
- Recall number
- Z-1859-2025
- Recalling firm
- Ultra Clean Systems, Inc.
- Product
- Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system
- Status
- Open, Classified
- Initiated
- 2025-03-31
- Posted
- 2025-05-28
- Terminated
- not on file
- Reason
- During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.
- Root cause
- Under Investigation by firm
- Action
- On 03/31/2025, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that as a result of an investigation, it has been revealed that a combination of chemical and electrical interactions that result in electrochemical degradation, compromising the structural integrity of the sensor over time. This degradation could potentially lead to water leakage into the machine. On 04/25/2025, the firm sent via FedEx and updated letter. Customers are notified and/or instructed that/to: -A field technicians will be contacting you to arrange for replacement if your unit has been impacted by this issue. -Complete and sign the attached response form to acknowledge that they have received the notification. Return the completed form to Getinge by emailing a scanned copy to recallresponses.qrc@getinge.com Public contact for questions/assistance - recalls.qrc@getinge.com
- Quantity
- 96 systems
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AZ, CA, CO, FL, ID, IL, IN, KS, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SD, TX, VA, WA, WI, and WY. The country of Canada.
- Lot, serial or model codes
- Triton 36 Model Number 6003000804 UDI-DI code: 00816634020013 Serial Numbers: 1601-XL36-249-1 1604-XL36-277-1 1808-XL36TR-596-1 1810-XL36TR-618-1 1905-XL36TR-707-1 1907-XL36TR-721-1 1911-XL36TR-800-1 2001-XL36TR-834-1 2004-XL36TR-859-1 2004-XL36TR-873-1 2104-XL36TR-947-1 2201-XL36TR-001-1 2203-XL36TR-017-1 2306-XL36TR-050-1 2306-XL36TR-051-1 2307-XL36TR-053-1 2312-XL36TR-120-1 2401-XL36TR-001-1 2402-XL36TR-017-1 2402-XL36TR-018-1 2402-XL36TR-027-1 2402-XL36TR-028-1 2402-XL36TR-029-1 2402-XL36TR-030-1 2405-XL36TR-048-1 2405-XL36TR-049-1 2405-XL36TR-050-1 2405-XL36TR-051-1 2405-XL36TR-060-1 2405-XL36TR-061-1 2405-XL36TR-062-1 2405-XL36TR-063-1 2406-XL36TR-064-1 2406-XL36TR-065-1 2406-XL36TR-072-1 2406-XL36TR-074-1 2406-XL36TR-075-1 2407-XL36TR-079-1 2407-XL36TR-083-1 2409-XL36TR-101-1 2409-XL36TR-103-1 2409-XL36TR-105-1 2409-XL36TR-107-1 2409-XL36TR-109-1 2409-XL36TR-110-1 2410-XL36TR-111-1 2410-XL36TR-113-1 2410-XL36TR-114-1 2411-XL36TR-118-1 2411-XL36TR-122-1 2412-XL36TR-135-1 XL36TR4381 XL36TR4961 XL36TR5111 2307-XL36TR-055-1 Triton 72 Model Number 6003000805 UDI-DI code: 00816634020006 Serial Numbers: 1511-XL72-230-1 1606-XL72-291-1 1706-XL72TR-401-1 1805-XL72TR-544-1 1805-XL72TR-546-1 1806-XL72TR-564-1 1806-XL72TR-565-1 1807-XL72TR-568-1 1809-XL72TR-611-1 1910-XL72TR-788-1 2002-XL72TR-844-1 2004-XL72TR-878-1 2004-XL72TR-879-1 2005-XL72TR-881-1 2006-XL72TR-890-1 2103-XL72TR-929-1 2103-XL72TR-930-1 2104-XL72TR-946-1 2112-XL72TR-1022-1 2202-XL72TR-013-1 2202-XL72TR-014-1 2202-XL72TR-015-1 2203-XL72TR-016-1 2207-XL72TR-045-1 2211-XL72TR-088-1 2304-XL72TR-028-1 2305-XL72TR-030-1 2305-XL72TR-034-1 2309-XL72TR-073-1 2310-XL72TR-099-1 2402-XL72TR-023-1 2402-XL72TR-024-1 2404-XL72TR-047-1 2405-XL72TR-052-1 2405-XL72TR-053-1 2406-XL72TR-073-1 2407-XL72TR-080-1 2407-XL72TR-084-1 2409-XL72TR-104-1 2411-XL72TR-120-1 XL72TR4401
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FLG
- Firm FEI
- 3002811406
- Firm city
- Oldsmar
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28