Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo Dynamics system is intended for the acceptance, transfer, display, storage, archive, and manipulation of digital medical images, including quantification and reportgeneration. syngo Dynamics is not intended to be used for reading mammography images.

Recall number
Z-1859-2013
Recalling firm
Siemens Medical Solutions USA, Inc
Product
syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo Dynamics system is intended for the acceptance, transfer, display, storage, archive, and manipulation of digital medical images, including quantification and reportgeneration. syngo Dynamics is not intended to be used for reading mammography images.
Status
Terminated
Initiated
2013-06-10
Posted
2013-08-01
Terminated
2015-02-24
Reason
Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the modality is configured at syngo Dynamics to end study at association close, and whe
Root cause
Software design
Action
Siemens sent a Customer Safety Advisory Notice dated June 4, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1) With sDIS implemented, use the DICOM Modality List (DMWL), without sDIS matching rules enabled OR 2) Do not implement sDIS. Update instructions were also provided. The reported issue will be resolved in a modification to syngo Dynamics version 9.5 and syngo Dynamics VA10A which will be released in the near future. We regret any inconvenience that this may cause , and we thank you in advance for your understanding. For further questions please call (610) 219-6300.
Quantity
9
Distribution
US Distribution only including the states of PA, OH, UT, CA, NY, MO, and TX.
Lot, serial or model codes
Model Number: 10091637, 10091805, with serial numbers 85164, 85289, 85296, 85297, 85149, 85116, 85177, 85051, and 85278.
510(k) numbers
["K102150"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register