FDA Medical Device Recalls (openFDA)
Nobel Biocare Usa Llc: Temporary Snap Abutment Engaging CC RP 1.5 mm; Pre-manufactured prosthetic components directly connected to the endosseous implant; catalog no. 38761 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.
- Recall number
- Z-1857-2020
- Recalling firm
- Nobel Biocare Usa Llc
- Product
- Temporary Snap Abutment Engaging CC RP 1.5 mm; Pre-manufactured prosthetic components directly connected to the endosseous implant; catalog no. 38761 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.
- Status
- Terminated
- Initiated
- 2019-08-09
- Posted
- not on file
- Terminated
- 2021-12-01
- Reason
- Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.
- Root cause
- Process control
- Action
- Urgent Medical Device Recall Letters were sent to U.S. customers via the U.S Postal Service and OUS customers via email beginning on August 9, 2019. This communication notified customers that the firm had identified a packaging issue affecting six product codes and a total of 7 lots of medical devices (titanium abutments) where the packages of abutment lots may contain an incorrect screw. Customers were informed that while the screw will tighten, it cannot be guaranteed that the screw will not loosen. If the screw loosens, customers are instructed to retighten the abutment with the correct screw. Customers were requested to: i1) identify all devices of the affected lots which they have in stock, 2) not to use the affected devices, 3) complete the Customer Acknowledgement Form and, 4) return it to Nobel Biocare via email at us.complaints@nobelbiocare.com or by fax to 714-282-5073, within 5 days of receipt of the Urgent Medical Device Recall Letter. Customers were instructed to return all devices from affected lots to Nobel Biocare, Attn: Returns PFA1914, 22715 Savi Ranch Parkway, Yorba Linda, CA 92887. The Urgent Medical Device Recall Letter notes that the manufacturer has reviewed the packaging process, has implemented corrective actions and will initiate preventive actions. The letter requests that customers pass the letter on to all those who need to be aware within their organizations or to any organization where the potentially affected devices have been transferred. Customers were instructed to contact their customer support representative at 1-800-322-5001, ext. 1527, if they require further information or support.
- Quantity
- 478 pieces
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AZ, AR, CA, CO, FL, GA, IL, IN, ID, KS, LA, MI, MD, ME, MA, MO, NY, NC, NJ, NE, NM, OH, OR, OK, PA, TX, TN, VA, WY, WA and WI. Countries of CA, CH, DE, ES, FR, GB, HU, IT, BR, CO, MX, CN, HK, IN, JP and TW.
- Lot, serial or model codes
- Catalog no.38761, Lot numbers 13077550 and 13077551; UDI: (01)07332747115604(10)13077550 and (01)07332747115604(10)13077551
- 510(k) numbers
- ["K161435"]
- PMA numbers
- not on file
- Product code
- NHA
- Firm FEI
- 1000125598
- Firm city
- Yorba Linda
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28