BARCO NV: Barco Medical Network Adapter, MNA-240, part of the NexxisOR system, Model Numbers: a) K9303078A, b) K9303097A
- Recall number
- Z-1854-2022
- Recalling firm
- BARCO NV
- Product
- Barco Medical Network Adapter, MNA-240, part of the NexxisOR system, Model Numbers: a) K9303078A, b) K9303097A
- Status
- Open, Classified
- Initiated
- 2022-07-27
- Posted
- 2022-09-29
- Terminated
- not on file
- Reason
- installation issues
- Root cause
- Labeling design
- Action
- Barco NV began notifying its consignees on about August 29, 2022, via email, about the device malfunction and stated that they would be providing an updated instructions for use (IFU). The firm followed with a Field Safety Action, via email on the same day, which described the issue, potential risks and a copy of the updated IFU. The letter requested the following: 1. Fill out and send back Response Form 1 to Barco Key Account Manager after receipt of this Notice. 2. Read the updated installation instructions, including the potential risk of wrong installation. 3. Inform relevant partners and ask them to evaluate if their installations must be adapted. 4. Fill out and send back Response Form 2 to Barco Key Account Manager after completion of the actions.
- Quantity
- 169 devices
- Distribution
- US Nationwide distribution in the states of FL, IL, NJ.
- Lot, serial or model codes
- a) K9303078A, UDI 05415334000056, all serial numbers; b) K9303097A, UDI 05415334001657, all serial numbers
- 510(k) numbers
- ["K173381"]
- PMA numbers
- not on file
- Product code
- DXJ
- Firm FEI
- 3013691682
- Firm city
- Kortrijk
- Firm state
- not on file
- Firm country
- Bulgaria
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28