Accumetrics Inc: VerifyNow System, part #85005-6H Product Usage: The VerifyNow P2Y12 assay device is a whole blood assay used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. The VerifyNow Systems intended for use with human whole blood and verify now test devices. The system should be operated by healthcare professionals trained on use of the system and in accordance with institution and policies and procedures. Accumetrics respresentatives will assist your institution in the installation of the system and operations of the operator.

Recall number
Z-1854-2012
Recalling firm
Accumetrics Inc
Product
VerifyNow System, part #85005-6H Product Usage: The VerifyNow P2Y12 assay device is a whole blood assay used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. The VerifyNow Systems intended for use with human whole blood and verify now test devices. The system should be operated by healthcare professionals trained on use of the system and in accordance with institution and policies and procedures. Accumetrics respresentatives will assist your institution in the installation of the system and operations of the operator.
Status
Terminated
Initiated
2012-02-06
Posted
2012-06-21
Terminated
2012-11-27
Reason
The recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and % Inhibition. The change implemented is the elimination of the BASE and % Inhibition results. There will be no change to the reporting of the PRU result whic
Root cause
No Marketing Application
Action
Accumetrics, Inc. sent an Urgent Field Correction notification letter and an Ackknowledgement - fax back form to all affected customers. The letter identifed the affected product, problem and actions to be taken. The letter informed customers that an Accumetrics representative will be contacting them to schedule an appointment to update their VerifyNow instrument software. The letter stated that Accumetrics realizes that procedures and report format may need to be revised as software changes is implemented, and will provide assistance to help in making this transition. For any questions or concerns, call at 1-800-643-1640 ext. 2; or email at: support@accumetrics.com.
Quantity
1,167 units total
Distribution
US Nationwide Distribution
Lot, serial or model codes
Not available
510(k) numbers
["K051231"]
PMA numbers
not on file
Product code
JOZ
Firm FEI
3001376288
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register