Skip to the content

FDA Medical Device Recalls (openFDA)

Anodyne Surgical: Hydrodissection Cannula (Chang), 27g, labeled as: anodynesurgical, REF 7018 ASICO, REF AS-7638 anodynesurgical, REF 7018NS katena, REF K20-3168 anodynesurgical, REF 104910 MSI, REF CA7800 ACCUTONE, REF AX14784-BULK Walcott Rx Products, REF RX2300NS Walcott Rx Products, REF RX2300

Recall number
Z-1852-2021
Recalling firm
Anodyne Surgical
Product
Hydrodissection Cannula (Chang), 27g, labeled as: anodynesurgical, REF 7018 ASICO, REF AS-7638 anodynesurgical, REF 7018NS katena, REF K20-3168 anodynesurgical, REF 104910 MSI, REF CA7800 ACCUTONE, REF AX14784-BULK Walcott Rx Products, REF RX2300NS Walcott Rx Products, REF RX2300
Status
Terminated
Initiated
2021-04-21
Posted
not on file
Terminated
2022-05-27
Reason
The product may have a yellowish-brown/greenish-brown residue present at the tip of the cannula.
Root cause
Employee error
Action
The firm initiated the recall by email on 04/21/2021. The letter explained the issue and requested the return of the product.
Quantity
8452 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of IL, NJ, TX, MO, CA, PA, WA, and the country of Canada.
Lot, serial or model codes
REF 7018, Lot Numbers: 201157-08 REF AS-7638, Lot Numbers: 210071-001, 210071-002 REF 7018NS, Lot Numbers: 210143-002 REF K20-3168, Lot Numbers:201108-005, 201108-017, 201108-018, 201154-007 REF 104910, Lot Numbers: 201144-002, 2010006-02, 210039-002, 210070-0001, 210080-01, 210127-001 REF CA7800, Lot Numbers: 201067-003, 201068-020 REF AX14784-BULK, Lot Numbers: 200580-006 REF RX2300NS, Lot Numbers: 200359-003 REF RX2300, Lot Numbers: 210142-002
510(k) numbers
not on file
PMA numbers
not on file
Product code
HMX
Firm FEI
1929756
Firm city
Saint Charles
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register