Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific Profile Single-Use Pediatric Snare 27mm UPN Outer Box: M00562571 Inner Pouch: M00562570 The single-use Profile Polypectomy Snare consists of a flexible wire cable loop which can be extended and retracted from the snare's flexible outer sheath using a three-ring handle. When passed through an endoscope and activated, the snare delivers a monopolar electrical current to cut and cauterize tissue with the loop

Recall number
Z-1852-2012
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Profile Single-Use Pediatric Snare 27mm UPN Outer Box: M00562571 Inner Pouch: M00562570 The single-use Profile Polypectomy Snare consists of a flexible wire cable loop which can be extended and retracted from the snare's flexible outer sheath using a three-ring handle. When passed through an endoscope and activated, the snare delivers a monopolar electrical current to cut and cauterize tissue with the loop
Status
Terminated
Initiated
2012-05-16
Posted
2012-06-21
Terminated
2013-02-28
Reason
Difficulty in extending snare loop from the catheter
Root cause
Process control
Action
Boston Scientific issued an Urgent Medical Device Recall Action letter via Federal Express Priority mail on May 16, 2012. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form. Scan and fax to 508-683-5578. For questions regarding this recall call 508-683-8000.
Quantity
1340 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Canada, Australia, Austria, Belgium, France, GB, Germany, Italy, Netherlands, Qatar, Spain, Sweden, and Switzerland
Lot, serial or model codes
Lot Number: 14974048
510(k) numbers
["K941750"]
PMA numbers
not on file
Product code
FDI
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register