Hospira Inc.: List No. 14200-28; SECONDARY BLOOD SET; 200 Micron Filter, 36 Inch, Non-DEHP; For Blood Product Administration; Made in Costa Rica; Product Inquiries should be directed to Hospira., Inc., Lake Forrest, IL 60045 USA Hospira Blood sets are designed for Blood Product Administration

Recall number
Z-1850-2013
Recalling firm
Hospira Inc.
Product
List No. 14200-28; SECONDARY BLOOD SET; 200 Micron Filter, 36 Inch, Non-DEHP; For Blood Product Administration; Made in Costa Rica; Product Inquiries should be directed to Hospira., Inc., Lake Forrest, IL 60045 USA Hospira Blood sets are designed for Blood Product Administration
Status
Terminated
Initiated
2013-04-01
Posted
2013-08-05
Terminated
2017-05-24
Reason
It was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.
Root cause
Device Design
Action
Hospira sent an Important Safety Information letter dated April 1, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to exercise EXTREME CAUTION when using Hospira blood bags and to make sure that all instructions for use are included with the blood bag and their facility's protocol for spiking blood bags are completely followed in order to minimize the possibility of puncturing the outer wall of the blood bag. Customers with issues were instructed to call Hospira at 1-800-441-4100 or email ProductComplaintsPP@hospira.com. Customers were asked to complete the attached replay formand return it to the fax number on the form. If customers distributed the product further they were instructed to notify their accounts that received the affected product and ask them to contact Stericycle at 1-866-608-3936 to receive a reply form. Customers with questions were instructed to contact Hospira at 1-800-441-4100. For questions regarding this recall call 224-212-2000.
Quantity
672 units
Distribution
Worldwide Distribution - USA (nationwide) and Puerto Rico and Internationally to Costa Rica.
Lot, serial or model codes
List Number: 14200-28;  Lot Numbers: 022045H, 022055H, 030605H, 201765H
510(k) numbers
["K101677"]
PMA numbers
not on file
Product code
BRZ
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register