Skip to the content

FDA Medical Device Recalls (openFDA)

Applied Medical Resources Corp: Applied Medical Dual Lumen Graft Cleaning Catheter & 1cc SYRINGE Nonpyrogenic, REF: A4GW6, 5F X 60cm, 9mm, Sterile EO, CE, UDI: (01)00607915111175

Recall number
Z-1848-2020
Recalling firm
Applied Medical Resources Corp
Product
Applied Medical Dual Lumen Graft Cleaning Catheter & 1cc SYRINGE Nonpyrogenic, REF: A4GW6, 5F X 60cm, 9mm, Sterile EO, CE, UDI: (01)00607915111175
Status
Terminated
Initiated
2019-10-24
Posted
2020-05-07
Terminated
2023-06-01
Reason
The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.
Root cause
Component change control
Action
On November 08, 2019, a "Urgent: Medical Device Recall" letter was sent via UPS to all affected customers. In addition to providing information on the recalled products, the letter asked customers to do the following: 1. Check your inventory for the recalled product. 2. Complete the Recall Notification Confirmation Form to acknowledge the recall and indicate if your facility is returning or has already used product from the lot listed above. 3. If no product is being returned, please indicate on the Recall Notification Confirmation Form. 4. If you are a distributor, please notify any facilities to which you distributed the affected product. Please also complete Page 4 of the Recall Notification Confirmation Form. 5. Return the Recall Notification Confirmation Form to Applied Medical by email to recall60810330@appliedmedical.com or faxing it to (949) 713-8908. 6. Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical (Product Return Instructions are on Page 5). 7. For product return questions, please contact Jaclene Rios-Simpson, Senior Manager, Customer Relations at (949) 713-8688 or jrios@appliedmedical.com. 8. For regulatory questions, please contact me, Lauren Contursi, at (949) 713-8767 or lcontursi@appliedmedical.com.
Quantity
28,799 catheters in total
Distribution
Worldwide distribution - United States: Nationwide OUS: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Poland, Saudi Arabia, South Africa, Spain, Taiwan, United Kingdom
Lot, serial or model codes
Model A4GW6, Applied Medical DUAL LUMEN GRAFT CLEANING CATETER, 9mm X 60cm, lot numbers 1267676, 1272322, 1284680, 1295323, 1306127, 1312586, 1330450, 1334826, 1338738, 1339555
510(k) numbers
["K973465"]
PMA numbers
not on file
Product code
DXE
Firm FEI
2027111
Firm city
Rancho Santa Margarita
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register