FDA Medical Device Recalls (openFDA)
Nellcor Puritan Bennett Inc. (dba Covidien LP): Puritan Bennett 840 Ventilator (4-840120DIUU-XX) The PB840 Ventilator System is intended to provide invasive or non-invasive ventilatory support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. It is used in hospitals and healthcare facilities.
- Recall number
- Z-1844-2014
- Recalling firm
- Nellcor Puritan Bennett Inc. (dba Covidien LP)
- Product
- Puritan Bennett 840 Ventilator (4-840120DIUU-XX) The PB840 Ventilator System is intended to provide invasive or non-invasive ventilatory support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. It is used in hospitals and healthcare facilities.
- Status
- Terminated
- Initiated
- 2014-06-03
- Posted
- 2014-06-20
- Terminated
- 2016-08-22
- Reason
- Covidien is conducting a medical device field correction for specific Puritan Bennett 840 Ventilators due to customer reports of loss of graphical user interface (GUI) display information while the ventilator continues to provide breath support.
- Root cause
- Nonconforming Material/Component
- Action
- Covidien sent an Urgent Medical Device Voluntary Field Correction letter dated 6/13/2014, to all affected customers. The letter identified the product the problem and and the action needed to be taken by the customer. Customers are requested to forward the recall letter to anyone they may have transferred or assigned for use one of the potentially affected ventilators. Service professionals will be identifying the affected PSUs and replacing them. The removed PSUs will be destroyed. To verify whether your devices are affected by this field safety corrective action and to initiate the process to receive the actions mentioned above, please complete the attached verification form and fax it to the Covidien contact indicated on the form. If you need further assistance, contact our Technical Support Department @ 1-800-255-6774, option 4, then option 1.
- Quantity
- 104
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY., and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Egypt, Germany, Guatemala, Israel, Japan, Kazakhstan, Kenya, Lebanon, Libya, Macedonia, Mexico, Montenegro, Morocco, Panama, Peru, Russia, Saudi Arabia, Singapore, Switzerland, Taiwan, Tunisia, Turkey, Yugoslavia (Serbia & Montenegro) Including GOVT/VA/Military
- Lot, serial or model codes
- serial number 3510100714 to 3512101160 and any units in which a replacement PSU was installed between March 2010 and September of 2010.
- 510(k) numbers
- ["K970460"]
- PMA numbers
- not on file
- Product code
- CBK
- Firm FEI
- 2936999
- Firm city
- Boulder
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28