Remel Inc: VersaTREK Instrument Series 240 REF 6240, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.

Recall number
Z-1841-2015
Recalling firm
Remel Inc
Product
VersaTREK Instrument Series 240 REF 6240, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.
Status
Terminated
Initiated
2015-05-19
Posted
2015-06-25
Terminated
2017-03-24
Reason
Use of the recalled product may result in false positive reports.
Root cause
Release of Material/Component prior to receiving test results
Action
Remel, Inc. a part of Thermo Scientific sent a Product Correction Notification dated May 19, 2015, to all affected customers. The notification states the problem with the device and that the recalling firm's Technical Services department will be contacting the customer to schedule a field service of the device to correct the issue. Customers were instructed to notify any personnel who need to be notified of the potential for a false positive resulting from rare inconsistent barometric pressure readings. Return the attached Medical Device Recall Return Response form acknowledging receipt of the notice of the affected product. If product was further distributed customers were instructed to contact those entities, advise them of the situation and provide them with a copy of the letter. Customers with questions were instructed to contact the Technical Services Department at 800-642-7029. For questions regarding this recall call 770-409-0713.
Quantity
143 devices
Distribution
Worldwide distribution: US (nationwide) to states of AZ, CO, FL, GA, IA, ID, LA, MI, MO, PR, TN, TX, VA, and WA and the countries of Canada, People's Republic of China, Egypt, Finland, France, Indonesia, Israel, Jordan, Kenya, Republic of Korea, Latvia, Romania, Russia, Spain, Thailand, United Kingdom, Uruguay, and Vietnam.
Lot, serial or model codes
Serial numbers: 0169039201140624, 0169040501140626, 0169050101141007, 0169068601141216, 0169078101150213, 2163X1013, 2164X1013, 2165X1013, 2166X1013, 2167X1013, 2168X1013, 2169X1013, 2170X1013, 2171X1013, 2172X1013, 2173X1013, 2174X1013, 2175X1013, 2176X1013, 2177X1013, 2178X1013, 2179X1013, 2180X1013, 2181X1013, 2182X1013, 2183X1013, 2184X1013, 2185X1013, 2186X1013, 2187X1013, 2188X1013, 2189X1013, 2190X1013, 2191X1013, 2192X1113, 2193X1113, 2194X1113, 2195X1113, 2196X1113, 2199X1113, 2201X1113, 2202X1113, 2203X1113, 2204X1113, 2205X1113, 2206X1113, 2207X1113, 2208X1113, 2209X1113, 2210X1113, 2211X1113, 2212X1113, 2213X1113, 2214X1113, 2216X1113, 2217X1113, 2218X1213, 2219X1213, 2220X1213, 2221X1213, 2222X1213, 2223X1213, 2224X1213, 2225X1213, 2226X1213, 2227X1213, 2228X1213, 2229X1213, 2230X1213, 2231X1213, 2232X1213, 2233X1213, 2234X1213, 2235X1213, 2237X1213, 2238X1213, 2239X1213, 2240X1213, 2241X1213, 2242X1213, 2243X1213, 2245X1213, 2249X0114, 2250X0114, 2251X0114, 2252X0114, 2253X0114, 2254X0114, 2255X0114, 2256X0114, 2257X0114, 2261X0114, 2262X0114, 2263X0114, 2264X0114, 2268X0114, 2269X0114, 2276X0214, 2277X0214, 2278X0214, 2279X0214, 2280X0214, 2281X0214, 2282X0214, 2286X0214, 2287X0214, 2288X0214, 2289X0214, 2290X0214, 2306X0714, 2307X0714, 2309X0714, 2311X0714, 2312X0714, 2314X0714, 2315X0714, 2316X0714, 2317X0714, 2318X0714, 2320X0714, 2321X0814, 2322X0814, 2323X0814, 2330X0814, 2332X0814, 2333X0814, 2334X0814, 2336X0814, 2337X0814, 2338X0814, 2339X0814, 2340X0814, 2342X0814, 2345X0914, 2346X0914, 2347X0914, 2348X0914, 2349X0914, 2352X0914, 2353X0914, 2359X0914, 2365X0914, and 2366X0914;
510(k) numbers
not on file
PMA numbers
not on file
Product code
MDB
Firm FEI
1924669
Firm city
Lenexa
Firm state
KS
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register