FDA Medical Device Recalls (openFDA)
Remel Inc: VersaTREK Instrument Series 240 REF 6240, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.
- Recall number
- Z-1841-2015
- Recalling firm
- Remel Inc
- Product
- VersaTREK Instrument Series 240 REF 6240, Manufactured for Remel Inc. The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile bodily fluids.
- Status
- Terminated
- Initiated
- 2015-05-19
- Posted
- 2015-06-25
- Terminated
- 2017-03-24
- Reason
- Use of the recalled product may result in false positive reports.
- Root cause
- Release of Material/Component prior to receiving test results
- Action
- Remel, Inc. a part of Thermo Scientific sent a Product Correction Notification dated May 19, 2015, to all affected customers. The notification states the problem with the device and that the recalling firm's Technical Services department will be contacting the customer to schedule a field service of the device to correct the issue. Customers were instructed to notify any personnel who need to be notified of the potential for a false positive resulting from rare inconsistent barometric pressure readings. Return the attached Medical Device Recall Return Response form acknowledging receipt of the notice of the affected product. If product was further distributed customers were instructed to contact those entities, advise them of the situation and provide them with a copy of the letter. Customers with questions were instructed to contact the Technical Services Department at 800-642-7029. For questions regarding this recall call 770-409-0713.
- Quantity
- 143 devices
- Distribution
- Worldwide distribution: US (nationwide) to states of AZ, CO, FL, GA, IA, ID, LA, MI, MO, PR, TN, TX, VA, and WA and the countries of Canada, People's Republic of China, Egypt, Finland, France, Indonesia, Israel, Jordan, Kenya, Republic of Korea, Latvia, Romania, Russia, Spain, Thailand, United Kingdom, Uruguay, and Vietnam.
- Lot, serial or model codes
- Serial numbers: 0169039201140624, 0169040501140626, 0169050101141007, 0169068601141216, 0169078101150213, 2163X1013, 2164X1013, 2165X1013, 2166X1013, 2167X1013, 2168X1013, 2169X1013, 2170X1013, 2171X1013, 2172X1013, 2173X1013, 2174X1013, 2175X1013, 2176X1013, 2177X1013, 2178X1013, 2179X1013, 2180X1013, 2181X1013, 2182X1013, 2183X1013, 2184X1013, 2185X1013, 2186X1013, 2187X1013, 2188X1013, 2189X1013, 2190X1013, 2191X1013, 2192X1113, 2193X1113, 2194X1113, 2195X1113, 2196X1113, 2199X1113, 2201X1113, 2202X1113, 2203X1113, 2204X1113, 2205X1113, 2206X1113, 2207X1113, 2208X1113, 2209X1113, 2210X1113, 2211X1113, 2212X1113, 2213X1113, 2214X1113, 2216X1113, 2217X1113, 2218X1213, 2219X1213, 2220X1213, 2221X1213, 2222X1213, 2223X1213, 2224X1213, 2225X1213, 2226X1213, 2227X1213, 2228X1213, 2229X1213, 2230X1213, 2231X1213, 2232X1213, 2233X1213, 2234X1213, 2235X1213, 2237X1213, 2238X1213, 2239X1213, 2240X1213, 2241X1213, 2242X1213, 2243X1213, 2245X1213, 2249X0114, 2250X0114, 2251X0114, 2252X0114, 2253X0114, 2254X0114, 2255X0114, 2256X0114, 2257X0114, 2261X0114, 2262X0114, 2263X0114, 2264X0114, 2268X0114, 2269X0114, 2276X0214, 2277X0214, 2278X0214, 2279X0214, 2280X0214, 2281X0214, 2282X0214, 2286X0214, 2287X0214, 2288X0214, 2289X0214, 2290X0214, 2306X0714, 2307X0714, 2309X0714, 2311X0714, 2312X0714, 2314X0714, 2315X0714, 2316X0714, 2317X0714, 2318X0714, 2320X0714, 2321X0814, 2322X0814, 2323X0814, 2330X0814, 2332X0814, 2333X0814, 2334X0814, 2336X0814, 2337X0814, 2338X0814, 2339X0814, 2340X0814, 2342X0814, 2345X0914, 2346X0914, 2347X0914, 2348X0914, 2349X0914, 2352X0914, 2353X0914, 2359X0914, 2365X0914, and 2366X0914;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MDB
- Firm FEI
- 1924669
- Firm city
- Lenexa
- Firm state
- KS
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28