FDA Medical Device Recalls (openFDA)
St. Jude Medical: Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain.
- Recall number
- Z-1836-2013
- Recalling firm
- St. Jude Medical
- Product
- Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain.
- Status
- Terminated
- Initiated
- 2012-07-26
- Posted
- 2013-07-31
- Terminated
- 2015-08-21
- Reason
- As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne
- Root cause
- Device Design
- Action
- St. Jude Medical sent a Important Medical Device Recall Advisory dated July 26, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To further assist in your patient care we are providing you with a list of all serial numbers we show have been distributed to you. ( See Attachment A ). Further questions please call (972) 309-2154.
- Quantity
- 34,617
- Distribution
- Worldwide Distribution--USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Belgium, Brazil, canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Saudi, South Africa, Spain, Sweden, Switzerland, Turkey and United Kingdom.
- Lot, serial or model codes
- Multiple lots recalled are the ones with supplier's inner battery lots of 326-508
- 510(k) numbers
- not on file
- PMA numbers
- ["P010032S023"]
- Product code
- LGW
- Firm FEI
- 1627487
- Firm city
- Plano
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28