Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499

Recall number
Z-1835-2020
Recalling firm
ICU Medical, Inc.
Product
ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499
Status
Terminated
Initiated
2019-07-12
Posted
not on file
Terminated
2023-05-10
Reason
Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary parameter.
Root cause
Under Investigation by firm
Action
Customers were notified of the recall via UPS letter on 7/12/2019. The firm is voluntarily notifying customers and will contact customers to schedule a software update once the updated software is available. The firm has instructed customers to use the mitigation instructions detailed in the customer letter to reduce the likelihood of the software conditions occurring. The firm also requested customers complete a response form and return it to the firm. For additional information or technical assistance contact ICU Medical at 1-866-829-9025, option 4.
Quantity
83 units
Distribution
US Nationwide Distribution in the states of FL, WI, NY, NC, VT, IL, MD, CA, NE, WA OUS: None
Lot, serial or model codes
All serial numbers of Cogent HMS, PN 58400-000, PN 58400-000R (User Interface Module (UIM), PN 58403-000 and Patient Interface Module (PIM), PN 58401-000) with Software Versions 1.3.2.460 and prior are impacted.
510(k) numbers
["K152006"]
PMA numbers
not on file
Product code
DXQ
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register