Skip to the content

FDA Medical Device Recalls (openFDA)

Signal Medical Corporation: Soft Face "No Bounce" Mallet Model/Catalog: BIO-1-1090

Recall number
Z-1834-2017
Recalling firm
Signal Medical Corporation
Product
Soft Face "No Bounce" Mallet Model/Catalog: BIO-1-1090
Status
Terminated
Initiated
2017-02-20
Posted
2017-04-17
Terminated
2017-08-25
Reason
Potential for poly tip of the mallet to come off during use. Potential for foreign material in the surgical field or delay in surgery. Reminder for customers that the poly tip must be removed for cleaning and sterilization to assure proper cleaning and sterilization of the entire mallet.
Root cause
Labeling design
Action
On 2/20/2016, URGENT: MEDICAL DEVICE RECALL notifications dated February 10, 2017 were sent to the affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers are to retain the notice and replace the old Instructions For Use with the updated version. Also, they are to acknowledge receipt of the instructions by signing the notice and emailing to SignalMedical2017@gmail.com or faxing to 810-364-7072. The firm has attached the revised labeling to the letter.
Quantity
1,640
Distribution
Nationwide Distribution.
Lot, serial or model codes
All units in distribution
510(k) numbers
not on file
PMA numbers
not on file
Product code
GFJ
Firm FEI
3002477077
Firm city
Marysville
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register