FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: Ultraflex Tracheobronchial Stent System. Uncovered - 16/60 Catalog Number: M00569240. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
- Recall number
- Z-1832-2010
- Recalling firm
- Boston Scientific Corporation
- Product
- Ultraflex Tracheobronchial Stent System. Uncovered - 16/60 Catalog Number: M00569240. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
- Status
- Terminated
- Initiated
- 2010-05-17
- Posted
- 2010-06-25
- Terminated
- 2011-10-20
- Reason
- Suture related device failures during deployment and repositioning of the stent
- Root cause
- Device Design
- Action
- Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - Only US distributed product is affected by recall
- Lot, serial or model codes
- Lot Numbers: 9788267 9788271 11049037 11056582 11056585 11403326 11403328 11431343 11717548 11755703 11830899 11831041 11951517 11951518 11959858 12073643 12147846 12150333 12150334 12351322 12396570 12570576 12622864 12659254 12833387 12898137 12946945 13199903
- 510(k) numbers
- ["K012883"]
- PMA numbers
- not on file
- Product code
- JCT
- Firm FEI
- 3005099803
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28