Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Ultraflex Tracheobronchial Stent System. Uncovered - 16/60 Catalog Number: M00569240. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.

Recall number
Z-1832-2010
Recalling firm
Boston Scientific Corporation
Product
Ultraflex Tracheobronchial Stent System. Uncovered - 16/60 Catalog Number: M00569240. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
Status
Terminated
Initiated
2010-05-17
Posted
2010-06-25
Terminated
2011-10-20
Reason
Suture related device failures during deployment and repositioning of the stent
Root cause
Device Design
Action
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Quantity
not on file
Distribution
Worldwide Distribution - Only US distributed product is affected by recall
Lot, serial or model codes
Lot Numbers: 9788267 9788271 11049037 11056582 11056585 11403326 11403328 11431343 11717548 11755703 11830899 11831041 11951517 11951518 11959858 12073643 12147846 12150333 12150334 12351322 12396570 12570576 12622864 12659254 12833387 12898137 12946945 13199903
510(k) numbers
["K012883"]
PMA numbers
not on file
Product code
JCT
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register