FDA Medical Device Recalls (openFDA)
Maquet Medical Systems USA: CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
- Recall number
- Z-1831-2024
- Recalling firm
- Maquet Medical Systems USA
- Product
- CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
- Status
- Open, Classified
- Initiated
- 2024-04-15
- Posted
- 2024-05-16
- Terminated
- not on file
- Reason
- The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
- Root cause
- Under Investigation by firm
- Action
- Consignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can also be contacted at 888-943-8872 (option 4, option 2, option 1) Monday through Friday between 8:00 AM and 6:00 PM EST to make service arrangements. Consignees are asked to not use affected devices for any additional patients until it has been cleared by the MCP/Getinge system check, repaired, or replaced. The provided recall notice is to be forwarded to users of the CARDIOHELP-i system and if the device was further distributed.
- Quantity
- 148 units (US: 5; OUS: 143)
- Distribution
- US Nationwide distribution in the states of NJ & TX.
- Lot, serial or model codes
- Part No. 70104.8012 and 70104.8012R; UDI-DI: 04037691658384; Serial no. 90410021, 90410029, 90410165, 90410254, 90410255, 90410024, 90410025, 90410026, 90410028, 90410030, 90410031, 90410034, 90410035, 90410036, 90410043, 90410044, 90410045, 90410046, 90410047, 90410048, 90410049, 90410050, 90410051, 90410052, 90410053, 90410054, 90410055, 90410056, 90410057, 90410058, 90410059, 90410060, 90410061, 90410062, 90410063, 90410064, 90410065, 90410067, 90410068, 90410069, 90410070, 90410071, 90410072, 90410073, 90410074, 90410075, 90410076, 90410077, 90410078, 90410079, 90410080, 90410081, 90410082, 90410084, 90410085, 90410086, 90410087, 90410088, 90410089, 90410091, 90410093, 90410094, 90410095, 90410096, 90410097, 90410098, 90410099, 90410100, 90410101, 90410102, 90410103, 90410104, 90410105, 90410106, 90410107, 90410108, 90410109, 90410110, 90410111, 90410112, 90410113, 90410114, 90410115, 90410116, 90410117, 90410118, 90410119, 90410120, 90410121, 90410122, 90410123, 90410124, 90410125, 90410126, 90410127, 90410130, 90410131, 90410132, 90410133, 90410134, 90410135, 90410136, 90410137, 90410138, 90410139, 90410140, 90410141, 90410142, 90410143, 90410144, 90410146, 90410147, 90410148, 90410149, 90410151, 90410152, 90410153, 90410154, 90410155, 90410156, 90410157, 90410158, 90410159, 90410160, 90410161, 90410162, 90410163, 90410166, 90410167, 90410168, 90410169, 90410170, 90410172, 90410173, 90410174, 90410175, 90410176, 90410177, 90410178, 90410179, 90410180, 90410181, 90410182, 90410183, 90410184, 90410185, 90410186, 90410187.
- 510(k) numbers
- ["K133598"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 3008355164
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28