Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Molecular Systems, Inc.: cobas EGFR Mutation Test, v2 and cobas cfDNA Sample Preparation Hungarian Translation Instructions for Use

Recall number
Z-1830-2016
Recalling firm
Roche Molecular Systems, Inc.
Product
cobas EGFR Mutation Test, v2 and cobas cfDNA Sample Preparation Hungarian Translation Instructions for Use
Status
Terminated
Initiated
2016-03-15
Posted
not on file
Terminated
2017-12-11
Reason
An error was found within the Hungarian translations of the cobas EGFR Mutation Test v2 Instructions for Use (M/N 07340761001-01HU, Doc Rev. 1.0, Dated 08/2015) and the cobas cfDNA Sample Preparation Kit Instructions for Use (M/N 07573758001-01HU, Doc. Rev. 1.0, Dated 05/2015).
Root cause
Packaging
Action
Roche Molecular Diagnostics Inc. issued an Urgent Field Safety Notice dated March 15, 2016 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Actions to be taken by the customer: " If you have utilized 2 L of DNA PBB during testing of plasma specimens, the samples must be retested using the correct volume (2mL DNA PBB). Communication of this Field Safety Corrective Action This notice must be passed on to all those who need to be aware within your organization or to any organization/individual where the potentially affected devices have been distributed/supplied. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. The undersigned confirms that this notice has been notified to the appropriate Regulatory Agency. For questions regarding this recall call 908-253-7200.
Quantity
8 kits
Distribution
Hungary
Lot, serial or model codes
EGFR v2: 07248563190 cfDNA: 07247737190
510(k) numbers
not on file
PMA numbers
not on file
Product code
OWD
Firm FEI
2243471
Firm city
S Branchburg
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register