Skip to the content

FDA Medical Device Recalls (openFDA)

Tosoh Smd Inc: AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.

Recall number
Z-1830-2013
Recalling firm
Tosoh Smd Inc
Product
AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.
Status
Terminated
Initiated
2013-05-15
Posted
2013-07-30
Terminated
2013-11-27
Reason
Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.
Root cause
Nonconforming Material/Component
Action
Tosoh Bioscience sent an Urgent Recall Notification letter dated May 15, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to respond on the amount of product that was destroyed by completing the attached form and faxing it to 614-317-1941. Customers with questions were instructed to call Tosoh Technical Support at 1-800-248-6764. For questions regarding this recall call 614-317-1909.
Quantity
11,309 bags
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. Internationally to Canada, Japan, Mexico, Chile, and Uruguay.
Lot, serial or model codes
The following lots numbers are affected by this recall:   060811, 070111, B11, B11-12/23/2011, B12-02/07/2012, B12-04/30/2012, B12-06/01/2012, B12-08/06/2012, B12-09/05/2012, B12-10/26/2012, and B12-11/27/2012.
510(k) numbers
["K971103"]
PMA numbers
not on file
Product code
KHO
Firm FEI
3005529799
Firm city
Grove City
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register