Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Ultraflex Tracheobronchial Stent System. Uncovered - 14/30 Catalog Number: M00569090. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.

Recall number
Z-1830-2010
Recalling firm
Boston Scientific Corporation
Product
Ultraflex Tracheobronchial Stent System. Uncovered - 14/30 Catalog Number: M00569090. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
Status
Terminated
Initiated
2010-05-17
Posted
2010-06-25
Terminated
2011-10-20
Reason
Suture related device failures during deployment and repositioning of the stent
Root cause
Device Design
Action
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Quantity
not on file
Distribution
Worldwide Distribution - Only US distributed product is affected by recall
Lot, serial or model codes
Lot Numbers: 9657941 9658105 9696978 9737028 9822915 9822918 11084897 11143496 11150354 11150355 11252315 11253322 11485085 11485088 11540266 11540268 11603795 11605485 11640294 11640296 11716316 11723616 11775490 11808563 11808564 11871267 11871268 12026243 12026244 12049248 12049249 12107672 12150138 12150139 12206312 12206316 12239697 12239880 12271577 12271579 12407609 12441821 12494755 12518886 12570571   12601507 12604799 12621237 12645436 12697020 12889006 12957090 12986896 13116725
510(k) numbers
["K012883"]
PMA numbers
not on file
Product code
JCT
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register