RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.): Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.

Recall number
Z-1829-2016
Recalling firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Product
Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.
Status
Terminated
Initiated
2016-04-22
Posted
2016-05-26
Terminated
2017-02-08
Reason
Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.
Root cause
Process design
Action
RTI sent an Urgent Voluntary Product Notification dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. RTI Surgical is conducting a voluntary recall of the Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve due to the potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip resulting in possible patient injury. For further questions, please call (906) 226-4489.
Quantity
17 devices
Distribution
US nationwide distribution in CA, FL, IN, MI, MO, NC, NY, and TX.
Lot, serial or model codes
Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve; Lot Number: 170684; Part Number: 26-COUNTER-T-XLINK
510(k) numbers
["K150254"]
PMA numbers
not on file
Product code
KWP
Firm FEI
1000115331
Firm city
Marquette
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register