Philips Healthcare Inc.: Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with FlatDetector with swivel cable Product Codes: 708034, 708037 This Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/ universal system. General R/F, performed along with more specialized interventional applications. This includes the following general areas: Digestive system Skeletal system Urinary system Reproductive system Respiratory system Circulatory system Various . Arthrograms . Myelograms . Facet joint injections . Discography SialographyFluoroscopy, Radiography and Angiography

Recall number
Z-1829-2012
Recalling firm
Philips Healthcare Inc.
Product
Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with FlatDetector with swivel cable Product Codes: 708034, 708037 This Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/ universal system. General R/F, performed along with more specialized interventional applications. This includes the following general areas: Digestive system Skeletal system Urinary system Reproductive system Respiratory system Circulatory system Various . Arthrograms . Myelograms . Facet joint injections . Discography SialographyFluoroscopy, Radiography and Angiography
Status
Terminated
Initiated
2012-05-21
Posted
2012-06-20
Terminated
2017-01-18
Reason
MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement
Root cause
Device Design
Action
Philips Healthcare Imaging issued Urgent Field Safety Notice dated May 21, 2012, to accounts informing an inspection will be performed of the unit. The Cable guiding plate will be replaced. If the cables show any damage the complete cable set wll also be replaced. No action is be taken by the customer. For further information customers should call 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Quantity
433 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Australia Austria Belgium Bharain Canada China Columbia Czech Republic Denmark Dutch Antilles Egypt Estonia Finland France Germany Hong Kong Ireland Israel Italy Latvia Lebanon Luxembourg Malaysia Mexico Netherlands New Zealand Norway Philippines Qatar Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syria Taiwan Thailand Turkey United Kingdom
Lot, serial or model codes
Units with cable swivel delivered from 2007
510(k) numbers
["K023441","K050151"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register