FDA Medical Device Recalls (openFDA)
Philips Healthcare Inc.: Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with FlatDetector with swivel cable Product Codes: 708034, 708037 This Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/ universal system. General R/F, performed along with more specialized interventional applications. This includes the following general areas: Digestive system Skeletal system Urinary system Reproductive system Respiratory system Circulatory system Various . Arthrograms . Myelograms . Facet joint injections . Discography SialographyFluoroscopy, Radiography and Angiography
- Recall number
- Z-1829-2012
- Recalling firm
- Philips Healthcare Inc.
- Product
- Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with FlatDetector with swivel cable Product Codes: 708034, 708037 This Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/ universal system. General R/F, performed along with more specialized interventional applications. This includes the following general areas: Digestive system Skeletal system Urinary system Reproductive system Respiratory system Circulatory system Various . Arthrograms . Myelograms . Facet joint injections . Discography SialographyFluoroscopy, Radiography and Angiography
- Status
- Terminated
- Initiated
- 2012-05-21
- Posted
- 2012-06-20
- Terminated
- 2017-01-18
- Reason
- MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement
- Root cause
- Device Design
- Action
- Philips Healthcare Imaging issued Urgent Field Safety Notice dated May 21, 2012, to accounts informing an inspection will be performed of the unit. The Cable guiding plate will be replaced. If the cables show any damage the complete cable set wll also be replaced. No action is be taken by the customer. For further information customers should call 1-800-722-9377. For questions regarding this recall call 978-687-1501.
- Quantity
- 433 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Australia Austria Belgium Bharain Canada China Columbia Czech Republic Denmark Dutch Antilles Egypt Estonia Finland France Germany Hong Kong Ireland Israel Italy Latvia Lebanon Luxembourg Malaysia Mexico Netherlands New Zealand Norway Philippines Qatar Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syria Taiwan Thailand Turkey United Kingdom
- Lot, serial or model codes
- Units with cable swivel delivered from 2007
- 510(k) numbers
- ["K023441","K050151"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28