Stryker Instruments Division of Stryker Corporation: Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker Instruments, Kalamazoo, MI. The DeKompressor Percutaneous Discectomy Probe is intended for use in the apiration of disc material during percutaneous discectomies in the lumbar, thoracic, and cervical regions of the spine
- Recall number
- Z-1829-2011
- Recalling firm
- Stryker Instruments Division of Stryker Corporation
- Product
- Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker Instruments, Kalamazoo, MI. The DeKompressor Percutaneous Discectomy Probe is intended for use in the apiration of disc material during percutaneous discectomies in the lumbar, thoracic, and cervical regions of the spine
- Status
- Terminated
- Initiated
- 2011-02-03
- Posted
- 2011-03-28
- Terminated
- 2011-06-01
- Reason
- The firm received five complaints that the stylet did not fit down the cannula. The investigation revealed that the incorrect probe assembly was packaged for two DeKompressor part numbers with the lot number of 10214012.
- Root cause
- Other
- Action
- Stryker Instruments Division of Stryker Corporation sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated February 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately check their inventory for any recalled product and quarantine any affected product. If the product was further distributed, customers were asked to forward the letter and the completed Reply Form to all affected locations. Customers were to be contacted to coordinate return of the affected product. Upon receipt of the product, a replacement would be issued to their account. For questions regarding this recall call 1-800-253-3210.
- Quantity
- 44 units both products
- Distribution
- Worldwide Distribution - USA including the states of PA, NY, KS, NJ, TX, MI, OK, and FL, and the countries of Italy and Switzerland.
- Lot, serial or model codes
- Lot: 10214012.
- 510(k) numbers
- ["K032473"]
- PMA numbers
- not on file
- Product code
- HRX
- Firm FEI
- 1811755
- Firm city
- Kalamazoo
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28