Integra LifeSciences Corp. (NeuroSciences): CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

Recall number
Z-1828-2026
Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Product
CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.
Status
Open, Classified
Initiated
2026-03-06
Posted
2026-04-17
Terminated
not on file
Reason
Software issue that renders the touch screen unresponsive.
Root cause
Under Investigation by firm
Action
An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice dated 3/6/26 was sent to consignees. The notice instructs consignees to restart systems when freeze to restore functionality of the touchscreen capabilities for 8 minutes; this action may be used multiple times. Integra's Service and Repair can provide remote support to disable and re-enable the write filter with Integra employees and/or Authorized service personnel. This will reset the console to provide approximately 28 days with no touchscreen issue. Integra Lifesciences Service Team will contact consignees within 5 to 10 business days to schedule correction of affected devices at no cost. Consignees with any questions can direct them to Customer Service by emailing custsvcnj@integralife.com or by phone at 1-800-654-2873.
Quantity
78 units
Distribution
Worldwide - US Nationwide distribution in the states of MI & NY. The countries of Belgium, China, Czechia, Dubai, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands. Poland, Spain, Switzerland, Tunisia & United Kingdom.
Lot, serial or model codes
Model: C7000; UDI: 10381780126232; Serial Numbers: CHA1902101IE, CHA2001301IE, CHA2001401IE, CHA2002301IE, CHA2202301IE, CHA2305101IE, CHA2305401IE, CHA2307001IE, CHA2402201IE, CHB1711001IE, CHB1723601IE, CHB2000801IE, CHB2100901IE, CHB2301401IE, CHB2500101IE, CHC1800601IE, CHC1902601IE, CHC1902701IE, CHC2001201IE, CHD1705801IE, CHD1707101IE, CHD1801001IE, CHD1801101IE, CHD1902501IE , CHE1707201IE, CHE1707301IE, CHE2203301IE, CHE2204001IE, CHE2301001IE , CHE2305001IE, CHE2502901IE, CHE2503101IE, CHE2503201IE, CHE2503401IE, CHE2503701IE, CHF1701401IE, CHF1702701IE, CHF2100701IE, CHF2303701IE, CHF2305401IE, CHG1700101IE, CHG1700801IE, CHG1902401IE, CHG2000201IE, CHG2000301IE, CHG2204101IE, CHG2402301IE, CHG2502801IE, CHG2506001IE, CHH1802901IE, CHH2402801IE, CHH2500601IE, CHJ2001101IE, CHJ2001301IE , CHJ2301201IE, CHJ2304601IE, CHK1700801IE, CHK1702301IE, CHK1703001IE, CHK2402601IE, CHK2403401IE, CHK2404101IE, CHL1701701IE, CHL1900201IE, CHL2102901IE , CHL2103001IE, CHL2301301IE, CHL2302001IE, CHL2302101IE, CHL2403301IE, CHM1901901IE, CHM2100101IE, CHM2101601IE, CHM2200201IE, CHM2200301IE, CHM2200501IE, CHM2300701IE, CHM2402101IE.
510(k) numbers
["K161882"]
PMA numbers
not on file
Product code
LFL
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register