Boston Scientific Corporation: Ultraflex Tracheobronchial Stent System. Uncovered - 10/30 Catalog Number: M00569070. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.

Recall number
Z-1828-2010
Recalling firm
Boston Scientific Corporation
Product
Ultraflex Tracheobronchial Stent System. Uncovered - 10/30 Catalog Number: M00569070. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.
Status
Terminated
Initiated
2010-05-17
Posted
2010-06-25
Terminated
2011-10-20
Reason
Suture related device failures during deployment and repositioning of the stent
Root cause
Device Design
Action
Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.
Quantity
not on file
Distribution
Worldwide Distribution - Only US distributed product is affected by recall
Lot, serial or model codes
Lot Numbers: 9752358 11001400 11001402 11057106 11057486 11133494 11134003 11206281 11206282 11261393 11261396 11337238 11337381 11444217 11444219 11499381 11499399 11543942 11543945 11592688 11592689 11664729 11664734 11688255 11688256 11724218 11724381 11794682 11794685 11845373 11845374 11914027 11914032 11979190 11979425 12034909 12034910 12122990 12173686 12173687 12185338 12185339 12314457 12396282 12535625   12586819 12635961 12706496 12749115 12833386 12882124 12937032 13056438 13084405 13198350
510(k) numbers
["K012883"]
PMA numbers
not on file
Product code
JCT
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register