Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America Llc: Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.

Recall number
Z-1827-2024
Recalling firm
Philips North America Llc
Product
Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.
Status
Open, Classified
Initiated
2024-04-19
Posted
2024-05-15
Terminated
not on file
Reason
Event Catalog information does not save when copied and transferred from one unit to another.
Root cause
Software design
Action
Consignees were sent an "URGENT Medical Device Correction," notification, dated 4/12/24. Consignees are instructed in the notification to review in Events Monitor and confirm the expected category is assigned for the reviewed event. If events are not assigned to the correct category consignees are not to depend on event notification for patient monitoring. The provided recall notification should be provided to all users of affected devices and to those the device was further distributed, for their awareness. Philips will contact consignees to schedule a visit from a Philips Field Service Engineer to provide a software patch (PIC iX 4.2.4) to correct the issue. Consignees with any questions are to contact their local Philips representative or call 1-800-722-9377.
Quantity
358 units (US: 45; OUS: 313)
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, MA, MI, MN, MO, NJ, NM, NY, OH, PA, TN, TX, WA, WV, & DC. The countries of AE, AT, AU, BE, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, IE, IL, IN, IT, JP, KR, LT, NL, NO, OM, PL, PT, RO, SE, SG, SK, TH, & ZA.
Lot, serial or model codes
Model No. 866389 and 866390. UDI-DI: n/a; Serial No. 7G6G-7PGV-W, 5971-2NL7-Z, 6D62-19MD-9, 463N-67ME-X, 6U1W-7TJA-9, 5K2A-6DHP-K, 5E42-35J4-P, 4E6Z-70HU-B, 1T12-0MNA-C, 0C0L-0FKE-4, 2H34-3MMP-U, 6P1E-7HJU-W, 3R6X-3XNY-6, 7F50-53NK-L, 3Z3B-1AHM-H, 0128-2WG5-0, 0K6E-39G4-5, 0C5Y-7FHX-V, 0X7R-2AGC-Y, 0P5D-5VMH-V, 2K0X-0TLA-F, 6627-6MKH-U, 5J7B-1XPT-T, 7U4C-44NB-E, 2Y7R-5KGZ-W, 1P22-0XNW-P, 1004-6RJX-E, 1U0P-0FKE-C, 396T-5ZH1-5, 0Y0Z-3GJ0-A, 7Y5Z-7FHX-F, 7U76-1HM0-D, 2X44-7ELE-J, 7Z3Y-0LJ3-2, 7R7A-6CPV-0, 3C44-7ELE-4, 2L0L-0FKE-A, 1V70-3XNY-9, 5A4T-57JE-R, 375F-7UMF-K, 6031-24GX-X, 2M2G-15LB-E, 2630-65P4-3, 6E64-7XLV-M, 3K33-61K2-0, 227N-7JG6-J, 6X1X-3CH6-V, 0T41-35J4-A, 4R0G-60JV-1, 414L-03P1-E, 6Y4B-44NB-J, 7C5C-7UMF-D, 1305-0UH7-M, 116V-21KC-Z, 4R1J-18LJ-4, 3A0H-6AMX-6, 4R03-1FHE-U, 5C4L-7WMK-2, 6B35-3MMP-T, 0X63-19MD-G, 350Y-2DMM-7, 2N37-04K8-1, 6Z1H-1KG1-6, 2P1T-75JN-B, 5X3N-67ME-P, 6K59-0HN3-B, 2V57-6TKK-N, 264N-2LKY-K, 0D1V-48PE-J, 6U32-24GX-0, 684V-57JE-Z, 610R-7AHH-2, 2K53-2END-0, 5G2W-1VNP-A, 5Y2X-65P4-E, 545X-32PN-F, 1R44-7ELE-2, 0R1W-2MNL-2, 2V51-29H4-H, 6T49-44NB-Z, 6B6G-3DKJ-4, 781X-48PE-W, 2M51-6KKD-E, 3055-68PX-3, 3V31-7NPM-F, 7Z05-0UH7-V, 0J71-6UGD-Z, 4R5J-5AGH-B, 116Z-1ZL8-V, 1K1B-1EGA-D, 1W6W-21KC-Z, 567W-7BG9-5, 0T1M-18LJ-R, 0F6V-6XMZ-5, 221G-5NMA-3, 4W6U-0PLF-2, 6F3G-46MR-K, 612A-2WG5-W, 1Z7W-00G0-F, 670F-60JV-1, 4T62-19MD-7, 523N-67ME-8, 067L-7MPL-2, 2J7X-2CJW-6, 017L-0VKT-B, 6633-24GX-D, 7676-1HM0-9, 4M77-1HM0-G, 3T5P-0YG8-Y, 003W-3UN7-M, 3D5U-2PJT-3, 117J-0BLD-V, 4K30-5BNT-7, 4V6C-1PJ5-R, 3T0R-7AHH-X, 325N-73HZ-G, 5Y3Z-2RHR-N, 4Y52-29H4-P, 5A4L-1CJJ-J, 724E-34MB-4, 4J6V-6XMZ-R, 156Y-70HU-R, 041V-48PE-D, 4N5A-1NH8-5, 6Z5N-0YG8-R, 2A5H-1YML-U, 5Y6J-3HL1-E, 6D2X-1BLC-4, 483K-28MC-M, 1036-4KHG-N, 6L0D-02P6-4, 5G1Y-1WJ2-5, 302H-6LPP-W, 5F6T-6XMZ-Y, 2N78-6CPV-9, 6Z6H-7PGV-1, 3F7R-56K0-H, 1J5H-1YML-3, 6R2J-47P7-9, 032M-79HL-V, 623Z-2RHR-9, 4W08-0UH7-U, 7A3M-6VGP-H, 5N40-2RHR-N, 002A-5MGJ-6, 4L7Y-00G0-F, 521J-18LJ-5, 121R-22HW-3, 1R7K-0VKT-L, 6M6N-0DN4-G, 472X-65P4-L, 3L6U-0PLF-V, 2C6N-3KLU-C, 4H5T-0YG8-P, 2A5D-1NH8-Y, 0M6B-7VM6-F, 0P4Y-78J7-V, 0A61-74NU-C, 4U39-3JGK-2, 5L5X-32PN-Z, 713Z-3UN7-Y, 7W3Y-0LJ3-B, 0Y5Z-7FHX-N, 6F2V-1BLC-W, 706P-0PLF-0, 737K-7LPC-G, 6U2G-3YNZ-J, 064C-44NB-H, 3F5L-5AGH-9, 6N1K-25G7-8, 6M5J-0JLM-H, 617X-31N5-0, 284W-72P0-G, 2K1C-42GL-Y, 0U28-2WG5-M, 1Y2E-0NNG-X, 0N1E-7HJU-5, 6L0F-60JV-U, 2V2W-3VM1-X, 1256-68PX-G, 6H5T-3FHA-4, 3K0J-6GLT-B, 591J-18LJ-G, 094B-40KU-3, 4U0A-3WPK-D, 0D6E-39G4-P, 2X7R-5KGZ-Z, 4V52-30PD-F, 0W65-4PJB-V, 115M-4LGG-4, 6N2W-7DL9-K, 2Z73-2NL7-2, 5K0P-17M9-H, 3107-4RG3-6, 2V3T-67ME-K, 5F72-2NL7-Y, 3B54-58M8-P, 1A4K-2BNM-C, 016X-52P8-4, 3153-2END-M, 0B5K-10GT-X, 3X12-2GGN-V, 0Z7K-0VKT-0, 5B0U-08NR-E, 2V03-1FHE-K, 4W2V-3VM1-7, 3Z3G-0KN0-0, 2Y5M-5AGH-E, 1D5N-0YG8-K, 0A3U-6ZJ1-5, 693T-6YM2-Y, 574E-40KU-P, 2X46-12JD-H, 4T7D-4EJH-E, 6Y79-1HM0-M, 0V7B-4EJH-A, 3Z4R-2LKY-9, 2141-35J4-G, 0V4G-0EG2-B, 4C76-1HM0-U, 2W26-0CKW-V, 5Y1Y-3CH6-A, 534U-57JE-Y, 7T0A-02P6-1, 4H2B-5YJP-0, 1J61-2KH5-M, 2L2P-6HHJ-D, 5G2A-6DHP-W, 7J0C-69LP-H, 1M6T-5ZH1-3, 6J3E-0KN0-2, 7A4Y-78J7-V, 6L3G-46MR-4, 4L7J-7MPL-J, 0Y0L-0FKE-E, 5H6A-2ZKV-1, 3F3D-41PZ-F, 661M-4HGE-D, 2P5H-0JLM-Y, 1135-04K8-Z, 7X0C-69LP-G, .
510(k) numbers
["K153702"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register