Philips Medical Systems, Inc.: Philips MR systems: Ingenia HA FLEX TRAK Trolley (for Ingenia 1.5T & 3.0T products) Product Usage: Diagnostic MR System. The Height Adjustable FlexTrak trolley and Trolley Variable Height IRF is an optional device designed to prepare the patient outside the examination room to enhance workflow and make additional non-magnetic hospital beds unnecessary.

Recall number
Z-1826-2014
Recalling firm
Philips Medical Systems, Inc.
Product
Philips MR systems: Ingenia HA FLEX TRAK Trolley (for Ingenia 1.5T & 3.0T products) Product Usage: Diagnostic MR System. The Height Adjustable FlexTrak trolley and Trolley Variable Height IRF is an optional device designed to prepare the patient outside the examination room to enhance workflow and make additional non-magnetic hospital beds unnecessary.
Status
Terminated
Initiated
2014-05-15
Posted
2014-06-18
Terminated
2015-08-07
Reason
Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.
Root cause
Device Design
Action
Philips sent an Urgent Field Safety Notice (FSN78100414) dated May 2014 via certified mail to US and Canadian consignees. Philips Key Market representatives will distribute the letter internationally. The letter identified the affected product, explains the problem, actions that should be taken in order to prevent risks for users, and the actions planned by Philips to correct the problem. On May 31, 2014, Philips released a Field Change Order that requires service engineers to replace the front and rear covers of the Ingenia MR FlexTrak Trolley with new covers with protection bumper that prevents foot clamping. For further information and support concerning this issue, consignees should contact their local Philips representative at 1-800-722-9377.
Quantity
420
Distribution
Worldwide Distribution - US (nationwide), Canada, Mexico Australia, Austria, Germany, Belarus, Bulgaria, Brazil, China, Denmark, Estonia, Finland, France, Germany, Greece, Italy, Jordan, Korea, Republic of Korea, Netherlands, New Zealand, Norway, Oman, Pakistan, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Lot, serial or model codes
Model Numbers: 9896 030 1929.2, 9896 030 1929.3
510(k) numbers
["K110151"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register