PHILIPS MEDICAL SYSTEMS: Philips Spectral CT on Rails. Model Number: 728334.

Recall number
Z-1825-2026
Recalling firm
PHILIPS MEDICAL SYSTEMS
Product
Philips Spectral CT on Rails. Model Number: 728334.
Status
Open, Classified
Initiated
2026-03-07
Posted
2026-04-16
Terminated
not on file
Reason
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? or should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Root cause
Under Investigation by firm
Action
Philips Medical notified consignees on about 03/07/2026 via letter. Consignees were instructed to identify any affected systems in inventory and specifically for each issue: 1. Issue #1: Ensure to follow the instructions mentioned in the IFU Section 4 Preparing for an Examination and Sub section Gantry Operation. 2. Issue #2: Avoid the Select Previous Surview Unavailable issue, manually remove the patient's middle name while importing the patient detail from HIS/RIS. 3. Issue #3: if Define Head Area option is grayed out with DoseRight Index (DRI) on, review the dose levels displayed on the User Interface, confirm and adjust the dose settings to appropriate levels before executing the scan as instructed in the IFU. Consignees were also instructed to complete and return the response form, to circulate the Urgent Medical Device Correction Letter to all users of the affected device so that they are aware of the issues, and retain the letter with affected system(s) until a solution is installed on each system; ensure the letter is in a place likely to be seen/viewed. Philips will contact each consignee to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and install the solution to resolve the issues (reference FCO72800830).
Quantity
3 systems
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
Lot, serial or model codes
Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
510(k) numbers
["K212875"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3017726341
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register