Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien LLC: EGIA 30 ARTICULATING MED THICK SULU, Item Code EGIA30AMT

Recall number
Z-1825-2019
Recalling firm
Covidien LLC
Product
EGIA 30 ARTICULATING MED THICK SULU, Item Code EGIA30AMT
Status
Open, Classified
Initiated
2019-05-17
Posted
not on file
Terminated
not on file
Reason
The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
Root cause
Process control
Action
All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology. " Complete the Recalled Product Return Form. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Quantity
3,113,280
Distribution
Nationwide domestic distribution, worldwide foreign distribution.
Lot, serial or model codes
lots N5A0871KX N7A0527KX N7E0441KX N8J0039KX N8J0527KX N8J0820KX
510(k) numbers
["K111825","K160176"]
PMA numbers
not on file
Product code
GDW
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register