Gentuity, LLC: Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;

Recall number
Z-1823-2026
Recalling firm
Gentuity, LLC
Product
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Status
Open, Classified
Initiated
2026-03-23
Posted
2026-04-16
Terminated
not on file
Reason
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
Root cause
Device Design
Action
On March 23, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: The customers will be required to follow the updated instructions to correct for repeated frames per the revised Operator Manual to address longitudinal measurement accuracy until the updated version of the software that addresses this issue is installed on the console within 6 months.
Quantity
184 units (28 US, 156 OUS)
Distribution
Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;
Lot, serial or model codes
Model Number: G10-01; UDI-DI: 00859910007032; Lot/Serial Number: 23C0113, 22J0101, 20A0202, 23F0101, 22F0113, 22L0100, 26B0203, 26B0200, 26B0204, 26B0202, 26B0201, 25L0302, 22E0204, 22J0102, 24H0101, 24H0102, 23K0201, 20A0203, 22K0306, 20A0201, 25J0200, 23C0110, 22K0305, 25E0100, 22F0110, 22K0304, 22K0302, 24B0107; OUS Lot numbers: 21G1900, 21G1901, 21G1902, 21G1903, 21G1904, 21H1101, 21H1102, 21H1103, 21H1104, 21H1105, 22C2300, 22C2301, 22C3000, 22D2600, 22D2700, 22D2701, 22D2702, 22D2705, 22E0500, 22E0501, 22F2700, 22F2701, 22F2702, 22F2703, 22F2704, 22G2600, 22G2601, 22G2700, 22H2900, 22H2901, 24B2100, 24B2101, 24B2102, 24B2103, 24B2104, 24C1301, 24C1302, 24C1303, 24C1304, 24C1305, 24D1504, 24D1505, 24D1506, 24D1507, 24D1508, 24E0800, 24E0801, 24E0802, 24E0803, 24E0804, 24F2408, 24F2409, 24F2410, 24F2411, 24F2412, 24G2205, 24G2206, 24G2207, 24G2208, 24G2209, 24H1400, 24H1401, 24H1402, 24H1403, 24H1404, 24M1000, 22L3000, 22L3001, 23A1904, PDA23-001, 23C2202, 23C2203, 23E0900, 23E0901, 23E0902, 23F2704, 23F2705, 23G2601, 23G2602, 23G2603, 23G2604, 23G0700, 23G0701, 23H3000, 23H3001, 23H3002, 23H3003, 23M1102, 23M1103, 23M1104, 23M1301, 24B2105, 24B2106, 24C2200, 24C2201, 24H1900, 24H1901, 24H1902, 24J0500, 24J0501, 24J0502, 24M1101, 24M1102, 24M1103, 24M1104, 24M1105, 25C0320, 25C0321, 25C0322, 25C0323, 25C0324, 25F2500, 25F2501, 25F2502, 25G3103, 25G3104, 25G3105, 25J0800, 25J0801, 25J0802, 25J2300, 25K0909, 25K0910, 25L0306, 25L0307, 25M0203, 25M0201, 25M0202, 26B0310, 26B0311, 26B2300, 26B1900, 26B1901, 26B1902;
510(k) numbers
["K242966"]
PMA numbers
not on file
Product code
DQO
Firm FEI
3011578544
Firm city
Sudbury
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register