FDA Medical Device Recalls (openFDA)
Gentuity, LLC: Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
- Recall number
- Z-1823-2026
- Recalling firm
- Gentuity, LLC
- Product
- Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
- Status
- Open, Classified
- Initiated
- 2026-03-23
- Posted
- 2026-04-16
- Terminated
- not on file
- Reason
- The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
- Root cause
- Device Design
- Action
- On March 23, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: The customers will be required to follow the updated instructions to correct for repeated frames per the revised Operator Manual to address longitudinal measurement accuracy until the updated version of the software that addresses this issue is installed on the console within 6 months.
- Quantity
- 184 units (28 US, 156 OUS)
- Distribution
- Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;
- Lot, serial or model codes
- Model Number: G10-01; UDI-DI: 00859910007032; Lot/Serial Number: 23C0113, 22J0101, 20A0202, 23F0101, 22F0113, 22L0100, 26B0203, 26B0200, 26B0204, 26B0202, 26B0201, 25L0302, 22E0204, 22J0102, 24H0101, 24H0102, 23K0201, 20A0203, 22K0306, 20A0201, 25J0200, 23C0110, 22K0305, 25E0100, 22F0110, 22K0304, 22K0302, 24B0107; OUS Lot numbers: 21G1900, 21G1901, 21G1902, 21G1903, 21G1904, 21H1101, 21H1102, 21H1103, 21H1104, 21H1105, 22C2300, 22C2301, 22C3000, 22D2600, 22D2700, 22D2701, 22D2702, 22D2705, 22E0500, 22E0501, 22F2700, 22F2701, 22F2702, 22F2703, 22F2704, 22G2600, 22G2601, 22G2700, 22H2900, 22H2901, 24B2100, 24B2101, 24B2102, 24B2103, 24B2104, 24C1301, 24C1302, 24C1303, 24C1304, 24C1305, 24D1504, 24D1505, 24D1506, 24D1507, 24D1508, 24E0800, 24E0801, 24E0802, 24E0803, 24E0804, 24F2408, 24F2409, 24F2410, 24F2411, 24F2412, 24G2205, 24G2206, 24G2207, 24G2208, 24G2209, 24H1400, 24H1401, 24H1402, 24H1403, 24H1404, 24M1000, 22L3000, 22L3001, 23A1904, PDA23-001, 23C2202, 23C2203, 23E0900, 23E0901, 23E0902, 23F2704, 23F2705, 23G2601, 23G2602, 23G2603, 23G2604, 23G0700, 23G0701, 23H3000, 23H3001, 23H3002, 23H3003, 23M1102, 23M1103, 23M1104, 23M1301, 24B2105, 24B2106, 24C2200, 24C2201, 24H1900, 24H1901, 24H1902, 24J0500, 24J0501, 24J0502, 24M1101, 24M1102, 24M1103, 24M1104, 24M1105, 25C0320, 25C0321, 25C0322, 25C0323, 25C0324, 25F2500, 25F2501, 25F2502, 25G3103, 25G3104, 25G3105, 25J0800, 25J0801, 25J0802, 25J2300, 25K0909, 25K0910, 25L0306, 25L0307, 25M0203, 25M0201, 25M0202, 26B0310, 26B0311, 26B2300, 26B1900, 26B1901, 26B1902;
- 510(k) numbers
- ["K242966"]
- PMA numbers
- not on file
- Product code
- DQO
- Firm FEI
- 3011578544
- Firm city
- Sudbury
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28