FDA Medical Device Recalls (openFDA)
U&I CORP.: Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.
- Recall number
- Z-1823-2020
- Recalling firm
- U&I CORP.
- Product
- Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.
- Status
- Terminated
- Initiated
- 2019-01-30
- Posted
- not on file
- Terminated
- 2020-11-20
- Reason
- The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.
- Root cause
- Employee error
- Action
- On 01/30/2019, "Notification for Sales Suspension and Product recall" and explanation of nonconformity notices were emailed to the domestic customer. The notification requests customers identify, isolate and return affected products immediately. It requests acknowledgement, and quantity of devices to be returned. The contact number for further questions is +82 31 860-6830 or kkjane@youic.com
- Quantity
- 42
- Distribution
- US Nationwide distribution in the state of CA.
- Lot, serial or model codes
- Lots: A18K220, A18L065, A18L013, A18L014, A18L015, A18L018, A18L019, A18L020, A18L021
- 510(k) numbers
- ["K172424"]
- PMA numbers
- not on file
- Product code
- OVE
- Firm FEI
- 3003200961
- Firm city
- not on file
- Firm state
- not on file
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28