Skip to the content

FDA Medical Device Recalls (openFDA)

U&I CORP.: Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.

Recall number
Z-1823-2020
Recalling firm
U&I CORP.
Product
Velofix SA cervical cage, REF: TSA161207, TSA161208, TSA161209, TSA161210, TSA161407, TSA161408, TSA161409, TSA161410 - Product Usage:The device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disk disease at one level from C2-C3 disc to the C7-T1 disc. Each product is composed of four components: PEEK cage, titanium plate, and two titanium markers.
Status
Terminated
Initiated
2019-01-30
Posted
not on file
Terminated
2020-11-20
Reason
The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less bone-screw engagement than intended by physicians.
Root cause
Employee error
Action
On 01/30/2019, "Notification for Sales Suspension and Product recall" and explanation of nonconformity notices were emailed to the domestic customer. The notification requests customers identify, isolate and return affected products immediately. It requests acknowledgement, and quantity of devices to be returned. The contact number for further questions is +82 31 860-6830 or kkjane@youic.com
Quantity
42
Distribution
US Nationwide distribution in the state of CA.
Lot, serial or model codes
Lots: A18K220, A18L065, A18L013, A18L014, A18L015, A18L018, A18L019, A18L020, A18L021
510(k) numbers
["K172424"]
PMA numbers
not on file
Product code
OVE
Firm FEI
3003200961
Firm city
not on file
Firm state
not on file
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28