Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Calcium Reagent, Calcium Test System for COBAS INTEGRA 400, 400 PLUS, 800, 800 CTS, cobas c 311, 501 and 502, Part number 20763128322, Roche Diagnostics Corporation, Indianapolis, IN. For the in vitro quantitative determination of calcium in serum, plasma and urine.

Recall number
Z-1823-2011
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Calcium Reagent, Calcium Test System for COBAS INTEGRA 400, 400 PLUS, 800, 800 CTS, cobas c 311, 501 and 502, Part number 20763128322, Roche Diagnostics Corporation, Indianapolis, IN. For the in vitro quantitative determination of calcium in serum, plasma and urine.
Status
Terminated
Initiated
2010-10-11
Posted
2011-03-25
Terminated
2011-09-13
Reason
The firm has recieved complaints of imprecision and erroneous patient results for calcium on the COBAS INTEGRA and COBAS c 501 system using reagent lots 62601901 and 63020401.
Root cause
Other
Action
The firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated October 11, 2010 to all consignees/customers via UPS(signature required) . The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the recalled product; discard from inventory according to their site's local regulations; complete and return the attached Faxback form via fax to 1-877-598-5705 even if they do not have the affected product, and file this letter for future reference. The firm added lot 63020401 to the recall in 3/2011. The firm is mailing an URGENT MEDICAL DEVICE REMOVAL notice to customers that purchased that lot of product. Please contact Roche Diagnostics Technical Support at 1-800-428-2336 if you have questions about the information contained in this letter.
Quantity
13, 405 units
Distribution
Nationwide distribution.
Lot, serial or model codes
Lots 62601901 and 63020401.
510(k) numbers
["K963292"]
PMA numbers
not on file
Product code
CIC
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register