FDA Medical Device Recalls (openFDA)
CooperSurgical, Inc.: Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295). The Mark I model is no longer produced, but the legacy humidifier bottle component design can also be used with the BT37 Mark II Benchtop Incubator.
- Recall number
- Z-1822-2022
- Recalling firm
- CooperSurgical, Inc.
- Product
- Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295). The Mark I model is no longer produced, but the legacy humidifier bottle component design can also be used with the BT37 Mark II Benchtop Incubator.
- Status
- Open, Classified
- Initiated
- 2022-08-12
- Posted
- 2022-09-28
- Terminated
- not on file
- Reason
- There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.
- Root cause
- Under Investigation by firm
- Action
- On August 12, 2022, the firm notified customers via an Urgent Medical Device Safety Notice. Customers were instructed to quarantine affected product in their inventory. Once customers complete and return the Acknowledgement and Receipt Form attached to the recall letter, the firm will arrange for the affected product to be returned at no cost. Customers will receive credit for returned affected product. CooperSurgical may be reached at 203-601-5200 ext. 03300 or recall@coopersurgical.com.
- Quantity
- 5,182 kits
- Distribution
- Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS MA MI MN MO NJ NY OH OR SC TX VA VT WA WI WV Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama.
- Lot, serial or model codes
- Lot Numbers 05593V291121 18-1192 1812-68 18-1268 G000575 G003811 G003893 G004517 G004874 G005109 G005476 G005723 E160415 Expiration date range: 27 Jun 2022 to 07 Dec 2024
- 510(k) numbers
- ["K121566"]
- PMA numbers
- not on file
- Product code
- MQG
- Firm FEI
- 1216677
- Firm city
- Trumbull
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as