FDA Medical Device Recalls (openFDA)
GE Healthcare: GE Healthcare Definium 8000 Digital Radiographic System. Formerly GE Healthcare Revolution XR/d with Tomosynthesis.
- Recall number
- Z-1822-2008
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Definium 8000 Digital Radiographic System. Formerly GE Healthcare Revolution XR/d with Tomosynthesis.
- Status
- Terminated
- Initiated
- 2008-03-07
- Posted
- 2008-09-25
- Terminated
- 2011-12-11
- Reason
- Loss of image data: Software anomaly in the Processing Software on the Definium 8000 system may impact patient safety when using the VolumeRAD advanced application (option). The slice visualization of VolumeRAD exams acquired on the Wallstand receptor will be offset by 17mm. This prevents the visualization of anatomy WITHIN a distance of 17mm from the wallstand patient barrier (receptor cover).
- Root cause
- Software design
- Action
- GE sent a Product Safety Notification, dated March 7, 2008, to all customers. The letter described the safety issue, products affected, and stated that a GE Healthcare Field Engineer will schedule a visit to each facility to install a software modification.
- Quantity
- 127
- Distribution
- Worldwide distribution: USA, Australia, Belgium, Canada, China, Finland, France, Germany, Great Britain, Italy, Netherlands, Spain, and Sweden.
- Lot, serial or model codes
- 00000002047M39, 00000005487M34, 00000096986DP%, 00000100740WK6, 00000TA41882-7, 00000TA41972-3, 00000TA41984-4, 00000TA41984-6, 00000TA42274-8, 00000TA42869-6, 00001004330WK5, 00001004332WK1, 00001004504WK5, 00001004505WK2, 00001004510WK2, 00001004513WK6, 00001004764WK5, 00001004765WK2, 00001004766WK0, 00001004768WK6, 00001004770WK2, 00001004772WK8, 00001005020WK1, 00001005022WK7, 00001005024WK3, 00001005030WK0, 00001005031WK8, 00001005035WK9, 00001005181WK1, 00001005182WK9, 00001005188WK6, 00001005260WK3, 00001005339WK5, 00001005758WK6, 00001005914WK5, 00001006042WK4, 00001006474WK9, 00001006635WK5, 00001006637WK1, 00001006638WK9, 00001006668WK6, 00001006871WK6, 00001007161WK1, 00001007166WK0, 00001007403WK7, 00001007404WK5, 00001007480WK5, 00001007536WK4, 00001007644WK6, 00001007650WK3, 00001007907WK7, 00001008312WK9, 00001008314WK5, 00001008497WK8, 00001008684WK1, 00001008807WK8, 00001008813WK6, 00001009140WK3, 00001009143WK7, 00001009145WK2, 00001009216WK1, 00001009268WK2, 00001009375WK5, 00001009380WK5, 00001009382WK1, 00001009709WK5, 00001009818WK4, 00001009875WK4, 00001009934WK9, 00001009976WK0, 00001010085WK7, 00001010099WK8, 00001010161WK6, 00001010239WK0, 00001010294WK5, 00001010295WK2, 00001010297WK8, 00001010431WK3, 00001010488WK3, 00001010524WK5, 00001010628WK4, 00001010629WK2, 00001010687WK0, 00001010693WK8, 00001010695WK3, 00001010906WK4, 00001010907WK2, 00001010964WK3, 00001011084WK9, 00001011085WK6, 00001011092WK2, 00001011215WK9, 00001011285WK2, 00001011289WK4, 00001011386WK8, 00001011455WK1, 00001011488WK2, 00001011490WK8, 00001011647WK3, 00001011654WK9, 00001011700WK0, 00001011712WK5, 00001011822WK2, 00001011825WK5, 00001011827WK1, 00001011828WK9, 00001011829WK7, 00001011832WK1, 00001012012WK9, 00001012119WK2, 00001012286WK9, 00001012449WK3, 00001012450WK1, 00001012453WK5, 00001012923WK7, 00001013064WK9, 00001013067WK2, 00001013100WK1, 00001013101WK9, 00001013134WK0, 00001013136WK5, 00001013223WK1, 00001013334WK6, 00001013336WK1, 00001013622WK4, 00001014310WK5, 00001014936WK7, 00001016147WK9
- 510(k) numbers
- ["K051967"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28