Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare: GE Healthcare Definium 8000 Digital Radiographic System. Formerly GE Healthcare Revolution XR/d with Tomosynthesis.

Recall number
Z-1822-2008
Recalling firm
GE Healthcare
Product
GE Healthcare Definium 8000 Digital Radiographic System. Formerly GE Healthcare Revolution XR/d with Tomosynthesis.
Status
Terminated
Initiated
2008-03-07
Posted
2008-09-25
Terminated
2011-12-11
Reason
Loss of image data: Software anomaly in the Processing Software on the Definium 8000 system may impact patient safety when using the VolumeRAD advanced application (option). The slice visualization of VolumeRAD exams acquired on the Wallstand receptor will be offset by 17mm. This prevents the visualization of anatomy WITHIN a distance of 17mm from the wallstand patient barrier (receptor cover).
Root cause
Software design
Action
GE sent a Product Safety Notification, dated March 7, 2008, to all customers. The letter described the safety issue, products affected, and stated that a GE Healthcare Field Engineer will schedule a visit to each facility to install a software modification.
Quantity
127
Distribution
Worldwide distribution: USA, Australia, Belgium, Canada, China, Finland, France, Germany, Great Britain, Italy, Netherlands, Spain, and Sweden.
Lot, serial or model codes
00000002047M39, 00000005487M34, 00000096986DP%, 00000100740WK6, 00000TA41882-7, 00000TA41972-3, 00000TA41984-4, 00000TA41984-6, 00000TA42274-8, 00000TA42869-6, 00001004330WK5, 00001004332WK1, 00001004504WK5, 00001004505WK2, 00001004510WK2, 00001004513WK6, 00001004764WK5, 00001004765WK2, 00001004766WK0, 00001004768WK6, 00001004770WK2, 00001004772WK8, 00001005020WK1, 00001005022WK7, 00001005024WK3, 00001005030WK0, 00001005031WK8, 00001005035WK9, 00001005181WK1, 00001005182WK9, 00001005188WK6, 00001005260WK3, 00001005339WK5, 00001005758WK6, 00001005914WK5, 00001006042WK4, 00001006474WK9, 00001006635WK5, 00001006637WK1, 00001006638WK9, 00001006668WK6, 00001006871WK6, 00001007161WK1, 00001007166WK0, 00001007403WK7, 00001007404WK5, 00001007480WK5, 00001007536WK4, 00001007644WK6, 00001007650WK3, 00001007907WK7, 00001008312WK9, 00001008314WK5, 00001008497WK8, 00001008684WK1, 00001008807WK8, 00001008813WK6, 00001009140WK3, 00001009143WK7, 00001009145WK2, 00001009216WK1, 00001009268WK2, 00001009375WK5, 00001009380WK5, 00001009382WK1, 00001009709WK5, 00001009818WK4, 00001009875WK4, 00001009934WK9, 00001009976WK0, 00001010085WK7, 00001010099WK8, 00001010161WK6, 00001010239WK0, 00001010294WK5, 00001010295WK2, 00001010297WK8, 00001010431WK3, 00001010488WK3, 00001010524WK5, 00001010628WK4, 00001010629WK2, 00001010687WK0, 00001010693WK8, 00001010695WK3, 00001010906WK4, 00001010907WK2, 00001010964WK3, 00001011084WK9, 00001011085WK6, 00001011092WK2, 00001011215WK9, 00001011285WK2, 00001011289WK4, 00001011386WK8, 00001011455WK1, 00001011488WK2, 00001011490WK8, 00001011647WK3, 00001011654WK9, 00001011700WK0, 00001011712WK5, 00001011822WK2, 00001011825WK5, 00001011827WK1, 00001011828WK9, 00001011829WK7, 00001011832WK1, 00001012012WK9, 00001012119WK2, 00001012286WK9, 00001012449WK3, 00001012450WK1, 00001012453WK5, 00001012923WK7, 00001013064WK9, 00001013067WK2, 00001013100WK1, 00001013101WK9, 00001013134WK0, 00001013136WK5, 00001013223WK1, 00001013334WK6, 00001013336WK1, 00001013622WK4, 00001014310WK5, 00001014936WK7, 00001016147WK9
510(k) numbers
["K051967"]
PMA numbers
not on file
Product code
KPR
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register