Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA HQ, Inc.: Synthes Synapse SD Drill Guide, Part Number SD389.477, Lot Number 6913900 Indicated for Hooks, Plate/Rods, Rods and Screws when intended to promote fusion of the cervical spine fusion of the cervical spine and occipitocervical junction (occiput-T3), the plate/rod, rod, hook and screw (3.2 mm cortex) components of the Synthes Cervifix; Axon and Synapse Systems are indicated for the following: -Degenerative Disk Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); -Spondylolisthesis; -Spinal Stenosis; -Fracture/dislocation; -Atlantoaxial fracture with instability; -Occipitocervical dislocation; -Revision of previous cervical spine surgery; and -Tumo

Recall number
Z-1821-2013
Recalling firm
Synthes USA HQ, Inc.
Product
Synthes Synapse SD Drill Guide, Part Number SD389.477, Lot Number 6913900 Indicated for Hooks, Plate/Rods, Rods and Screws when intended to promote fusion of the cervical spine fusion of the cervical spine and occipitocervical junction (occiput-T3), the plate/rod, rod, hook and screw (3.2 mm cortex) components of the Synthes Cervifix; Axon and Synapse Systems are indicated for the following: -Degenerative Disk Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); -Spondylolisthesis; -Spinal Stenosis; -Fracture/dislocation; -Atlantoaxial fracture with instability; -Occipitocervical dislocation; -Revision of previous cervical spine surgery; and -Tumo
Status
Terminated
Initiated
2012-10-12
Posted
2013-07-29
Terminated
2015-06-11
Reason
A complaint was received from the field, that the drill guide was etched incorrectly, 12mm etch instead of 18mm drill guide. The firm identified that this constituted a field action and retrospectively reported this action.
Root cause
Labeling Change Control
Action
The firm, Synthes USA, sent an email to its customer. The email identified the product, problem and the actions to be taken. The customer was instructed to do the following: do not use this product; do not remove any product from your account or return any product without further instructions. If you have identified that you have the product, please have the product located in an area that ensures it will not be used. Further instructions will be sent out with regards to any action that needs to be taken next. Please call Synthes if you have any questions at 610-719-5364.
Quantity
2
Distribution
US distribution: PA only. .
Lot, serial or model codes
Part Number SD389.477, Lot Number 6913900
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register