FDA Medical Device Recalls (openFDA)
Biomet, Inc.: Various trauma and sports medicine instruments and implants. Rod, fixation, intramedullary and accessories.
- Recall number
- Z-1819-2016
- Recalling firm
- Biomet, Inc.
- Product
- Various trauma and sports medicine instruments and implants. Rod, fixation, intramedullary and accessories.
- Status
- Terminated
- Initiated
- 2016-04-01
- Posted
- 2016-05-26
- Terminated
- 2017-05-23
- Reason
- Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
- Root cause
- Equipment maintenance
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
- Lot, serial or model codes
- Part #'s 28234 28338 28811 28813 28821 29260 29822 805065080 14-400428 14-400436 14-405032 14-405040 14-440118 14-440121 14-441280 14-440066S Lot #'s 590970 591080 591090 665210 692040 692250 692280 692290 695900 610470 665230 665240 692440 692450 692480 692620 692660 692670 692690 721680 721720 721730 729080 722730 621840 621860 678300 664740 664820 664870 664880 503760 503770 693840 476230
- 510(k) numbers
- ["K050118","K063570","K091976","K871539"]
- PMA numbers
- not on file
- Product code
- HSB
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as