Skip to the content

FDA Medical Device Recalls (openFDA)

DICOM Grid, Inc. d/b/a Ambra Health: Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and analysis tool for diagnostic images.

Recall number
Z-1818-2022
Recalling firm
DICOM Grid, Inc. d/b/a Ambra Health
Product
Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and analysis tool for diagnostic images.
Status
Open, Classified
Initiated
2022-06-23
Posted
2022-09-27
Terminated
not on file
Reason
A race condition between the storage system and services database has the potential to revert edits made to patient information upon first ingestion of a study.
Root cause
Software design
Action
The firm distributed an Urgent Medical Device Recall Notice dated June 23, 2022, by email to affected customers. The firm informed customers that a race condition was identified between Ambra's storage system and the serivces database which has the potential to revert edits made to patient health information upon first ingestion of a study. This could lead to a possibility that incorrect patient ifnormation is presented to the physician, which would impact the decision making in regard to patient care. Customers are asked to advise all users of the system of this issue, request a patch from Ambra to be enabled on your system, and wait until all images for a given study are ingested before manually editing patient health information in the study. Ambra's actions include the deployment of a patch that is acting as a temporary fix until a software update can be deployed to all customers, which will prevent the issue from occurring in the future.
Quantity
671 downloads
Distribution
Domestic: AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, DC, & PR.
Lot, serial or model codes
UDI-DI: AMBRHEALTHSOLUTIONS0; Software Version: 3.22.2.0 and 3.22.3.0.
510(k) numbers
["K202335"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3008776294
Firm city
New York
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register