FDA Medical Device Recalls (openFDA)
Biomet, Inc.: Various trauma and sports medicine instruments and implants. Plate, fixation, bone.
- Recall number
- Z-1817-2016
- Recalling firm
- Biomet, Inc.
- Product
- Various trauma and sports medicine instruments and implants. Plate, fixation, bone.
- Status
- Terminated
- Initiated
- 2016-04-01
- Posted
- 2016-05-26
- Terminated
- 2017-05-23
- Reason
- Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
- Root cause
- Equipment maintenance
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
- Quantity
- not on file
- Distribution
- Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
- Lot, serial or model codes
- Part #'s 24616 131212112 131212113 131212114 131212116 131212118 131212216 131212218 131212220 131212222 131212224 131212316 131212318 131212320 131212322 856135016 856135018 856135020 856135022 856135024 856135026 856135028 856135030 856135034 856135036 856135038 856135040 856135042 Lot #'s 606160 729130 664900 752650 752660 664910 664930 752680 664950 752690 752720 664990 665020 665040 665060 665070 665090 665100 665120 665130 758430 758500 758520 758540 758550 758570 758670 758680 758700 758710 758720 758730 758740
- 510(k) numbers
- ["K093880","K111663"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28