WELCH ALLYN, INC/MORTARA: S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor provides a means to acquire and transmit simultaneous ECG and SpO2 data to a Surveyor Central Station monitoring system while allowing the patient to be ambulatory within the range of the WiFi network.

Recall number
Z-1815-2020
Recalling firm
WELCH ALLYN, INC/MORTARA
Product
S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor provides a means to acquire and transmit simultaneous ECG and SpO2 data to a Surveyor Central Station monitoring system while allowing the patient to be ambulatory within the range of the WiFi network.
Status
Terminated
Initiated
2020-03-18
Posted
2020-04-29
Terminated
2026-02-09
Reason
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Root cause
Device Design
Action
Welch Allyn notified customers on 2/19/2020 via "Voluntary Medical Device Correction Welch Allyn Patient Cables or Lead Sets". The recall letter identified the affected product and stated the reason for the recall. The firm also requested the distributors to immediately notify accounts if the product was further distributed.
Quantity
78,521 units total
Distribution
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Lot, serial or model codes
Model Numbers: S4-Q-ASX-XXX, S4-Q-AXX-XXX, S4-P-A, 9293-061-50, 9293-061-51, 9293-061-52, 9293-061-53.
510(k) numbers
["K160685"]
PMA numbers
not on file
Product code
DPS
Firm FEI
2183461
Firm city
Milwaukee
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register